ASL and Sodium MRIMRI Physics

Arterial spin labelling in plain English

Arterial spin labelling (ASL) measures cerebral blood flow without an injected contrast agent. Instead it uses the patient’s own blood as a tracer. This article explains the idea, the main variants and what the numbers do and do not tell you.

Learning objectives

  • Explain how magnetically labelled blood works as an endogenous tracer.
  • Describe the label and control images and why their difference reflects perfusion.
  • Define labelling duration and post-labelling delay and why they matter.
  • List the main ASL labelling approaches and typical limitations.
Arterial spin labelling: labelled and control images and the resulting perfusion image
Arterial spin labelling: labelled and control images and the resulting perfusion image. Image: Alpheratzstar, Public domain, via Wikimedia Commons (File:Arterial_Spin_Labeling-experiment.png).

The core idea

Blood flowing towards the brain is magnetically labelled, typically in the neck. After a short wait, the labelled blood has reached the brain tissue and slightly changes the signal. A second image is acquired with no labelling (the control). Subtracting label from control cancels the static tissue and leaves a small signal proportional to blood delivered.

The difference signal is only a small percentage of the total, so ASL typically averages many label-control pairs to improve signal-to-noise ratio.

Timing: labelling duration and post-labelling delay

Two timing parameters shape the signal. Labelling duration is how long blood is labelled. Post-labelling delay (PLD) is the wait between the end of labelling and image acquisition, which lets labelled blood arrive in tissue. Too short a PLD leaves labelled blood in the arteries; too long loses signal as the label decays with T1.

Timing diagram showing a labelling period, a post-labelling delay and then image acquisition
ASL timing: labelling duration, post-labelling delay and readout. Schematic only; real values depend on the protocol.

Main labelling approaches

ApproachHow labelling is doneStrengthsLimitations
Pulsed ASL (PASL)Short pulse labels a slab of bloodEfficient labellingLower signal-to-noise ratio in many implementations
Continuous ASL (CASL)Long continuous labellingHigh labelling effectNeeds special hardware; high energy deposition
Pseudo-continuous ASL (PCASL)Train of pulses mimics continuous labellingGood balance of signal and practicality on standard scannersSensitive to off-resonance in the labelling plane

A consensus paper from the ISMRM Perfusion Study Group and the European ASL in Dementia consortium recommends PCASL with a 3D readout for clinical use (Alsop et al., 2015).

What ASL numbers mean

With appropriate modelling and calibration, ASL can be converted to cerebral blood flow in mL/100 g/min. The numbers depend on assumptions such as arterial transit time, labelling efficiency and blood T1. Treat quantitative values as protocol-dependent estimates and compare within the same method.

Limitations and artefacts

  • Low signal-to-noise ratio compared with contrast-based methods.
  • Prolonged arterial transit time (for example in slow flow) can mimic low perfusion.
  • Motion between label and control images corrupts the subtraction.
  • Off-resonance and susceptibility effects reduce labelling efficiency.

Key takeaways

  • ASL uses magnetically labelled blood as a tracer, no contrast injection.
  • Label minus control gives a perfusion-weighted signal.
  • Labelling duration and post-labelling delay are critical choices.
  • PCASL with 3D readout is the consensus recommendation for clinical use.

Frequently asked questions

Does ASL need a contrast agent?

No. The labelled blood acts as the tracer.

Why is the image noisy?

The perfusion signal is a small fraction of the image signal, so many repetitions are averaged.

Can I compare absolute values between sites?

Only with caution. Different labelling, timing and modelling choices affect the result.

References and further reading

  • Alsop DC, et al. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications: a consensus of the ISMRM Perfusion Study Group and the European Consortium for ASL in Dementia. Magn Reson Med. 2015;73:102-116.

Related articles

Last reviewed: 4 November 2026. Educational content for researchers, clinicians and students. It does not replace local protocols, manufacturer guidance or clinical judgement.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles

Leave a Reply

Your email address will not be published. Required fields are marked *

Back to top button