Larmor frequency explained: why MRI works at radio frequencies
Every MRI scanner talks to the body in radio frequencies, and the number it uses is set by one equation. The Larmor frequency tells you how fast nuclear spins precess in a magnetic field, and therefore which radio frequency can excite them. This article explains the idea from first principles and connects it to real scanners.
Learning objectives
- State the Larmor equation and define each term.
- Calculate the resonance frequency of hydrogen at common field strengths.
- Explain why different nuclei need different frequencies at the same field.
- Understand why field strength influences RF hardware and why ultra-high-field work uses dedicated coils.

Spins, fields and precession
Hydrogen nuclei (protons) have a property called spin, which gives each one a tiny magnetic moment. In a strong static magnetic field B₀, these moments do not simply line up like compass needles. They precess around the field direction, in the same way a tilted spinning top wobbles around the vertical. The rate of that precession is the Larmor frequency.
The Larmor equation
f = (γ / 2π) × B₀
f = Larmor frequency (MHz)
γ/2π = gyromagnetic ratio of the nucleus (MHz per tesla)
B₀ = static magnetic field strength (tesla)
The frequency is directly proportional to field strength. For hydrogen, γ/2π is about 42.58 MHz per tesla, so doubling the field doubles the frequency.
Worked examples for hydrogen
| Field strength | Calculation | Larmor frequency |
|---|---|---|
| 1.5 T | 42.577 × 1.5 | ≈ 63.9 MHz |
| 3 T | 42.577 × 3 | ≈ 127.7 MHz |
| 7 T | 42.577 × 7 | ≈ 298.0 MHz |
These frequencies lie in the radio-frequency band, which is why an MRI scanner transmits and receives with RF coils rather than anything resembling light or X-rays.
Why resonance matters
To tip the magnetisation away from B₀ and create a measurable signal, the scanner applies a brief RF pulse at the Larmor frequency. Only RF energy at (or very near) this frequency interacts efficiently with the spins. This is resonance. After the pulse, the precessing magnetisation induces a voltage in the receive coil, and that is the MR signal.
Different nuclei, different frequencies
Each nucleus has its own gyromagnetic ratio, so at the same field it resonates at a different frequency. This is why sodium or phosphorus imaging needs the scanner and coils to be tuned to a frequency other than the hydrogen one. At 7 T, hydrogen resonates near 298 MHz while sodium-23 resonates near 79 MHz.
| Nucleus | γ/2π (MHz/T, approx.) | At 3 T (MHz) | At 7 T (MHz) |
|---|---|---|---|
| ¹H | 42.577 | 127.7 | 298.0 |
| ²³Na | 11.262 | 33.8 | 78.8 |
| ³¹P | 17.235 | 51.7 | 120.6 |
| ¹³C | 10.708 | 32.1 | 75.0 |
| ¹⁹F | 40.052 | 120.2 | 280.4 |
You can explore other values with the Larmor frequency calculator.
Small shifts: chemical environment
The frequency a nucleus actually experiences depends very slightly on its surroundings, because electrons nearby shield part of the field. These shifts are measured in parts per million (ppm) and are the basis of MR spectroscopy. They are tiny compared with the Larmor frequency itself, but they let us tell molecules apart.
Practical consequences
- RF hardware must be designed for the Larmor frequency of each nucleus and field strength.
- Wavelength effects become more important as frequency rises, which is one reason ultra-high-field imaging needs careful RF design.
- Field drift or inhomogeneity changes the local frequency, which can cause artefacts or poor excitation.
Key takeaways
- Larmor frequency is proportional to B₀ and set by the nucleus.
- Hydrogen at 3 T resonates near 127.7 MHz; at 7 T near 298 MHz.
- RF pulses must match this frequency to excite the spins.
- Other nuclei, such as sodium, require different tuning at the same field.
Frequently asked questions
Is the Larmor frequency the same as the scanner frequency?
The scanner works at a centre frequency close to the Larmor frequency, and the exact value is set during calibration for each patient and scan.
Why not use higher fields for everything?
Higher fields increase signal but also bring practical challenges such as RF wavelength effects, energy deposition limits and susceptibility artefacts. The best field depends on the question.
Are the constants in this article exact?
They are rounded values suitable for teaching. Use scanner-reported frequencies for real work.
References and further reading
- Bloch F. Nuclear induction. Phys Rev. 1946;70:460-474.
- Brown RW, Cheng YCN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design. 2nd ed. Wiley; 2014.
Related
- Larmor frequency calculator
- Arterial spin labelling in plain English
- Next: T1 and T2 relaxation explained