Sodium (²³Na) MRI explained: what it measures and why it is hard

Hydrogen gives MRI its familiar anatomical detail, but other nuclei carry information too. Sodium-23 is one of the most interesting: sodium concentration is tightly regulated in healthy cells, so changes in it can signal tissue damage or altered metabolism. This article explains what sodium MRI measures, why it is technically demanding, and how it is done.

Learning objectives

  • Explain why sodium is of biological interest.
  • Describe why sodium MRI signal is much weaker than proton signal.
  • Explain why ultra-short echo times and special readouts are used.
  • Understand what “total sodium concentration” means and how it is estimated.
Whole-body sodium (23Na) MRI of a volunteer at 3 T
Whole-body sodium (23Na) MRI of a volunteer at 3 T. Image: SodiumMRI, CC BY-SA 4.0, via Wikimedia Commons (File:Whole-body_sodium_magnetic_resonance_images_of_female_human.png).

Why sodium?

Sodium concentration differs sharply between compartments: it is much higher outside cells than inside. Cells spend energy maintaining this gradient. When energy supply or membrane function is disturbed, intracellular sodium rises, and tissue sodium concentration may change. Sodium MRI offers a non-invasive route to this information, which is why it is explored in stroke, tumours, multiple sclerosis, neurodegeneration and epilepsy research.

Sodium compared with hydrogen

Property¹H (hydrogen)²³Na (sodium)
Larmor frequency at 3 T≈ 127.7 MHz≈ 33.8 MHz
Larmor frequency at 7 T≈ 298.0 MHz≈ 78.8 MHz
Tissue concentrationVery high (water, fat)Far lower
Relative MR sensitivityReferenceLower
RelaxationRelatively slowVery fast, often described with two T2 components

Frequencies follow from the Larmor equation (see Larmor frequency explained). Overall, the sodium signal in tissue is several thousand times weaker than the proton signal, so sodium images are low in resolution compared with proton images and need longer averaging.

Fast, two-component relaxation

Sodium-23 has spin 3/2 and a quadrupole moment, which interacts with local electric field gradients. In many tissues this produces biexponential transverse relaxation: a very short component and a longer one, with time constants in the order of a few milliseconds and tens of milliseconds respectively. The practical consequence is that the signal decays very rapidly after excitation, and conventional sequences with echo times of several milliseconds lose most of it.

How sodium images are acquired

  • Ultra-short echo time (UTE) readouts, such as 3D radial or twisted-projection trajectories, begin sampling almost immediately after excitation.
  • Averaging of many acquisitions recovers signal-to-noise ratio, so scans take longer than a typical proton scan.
  • Dedicated RF coils tuned to the sodium frequency, often dual-tuned with a proton channel, are required.
  • Higher field strengths such as 7 T improve signal-to-noise ratio, which is one reason ultra-high-field research is active in this area.

Total sodium concentration

To report a number in mmol/L, the sodium signal is usually calibrated against reference vials of known sodium concentration placed in the field of view. Corrections for coil sensitivity and relaxation are needed. The result is often called total sodium concentration (TSC). It is an average over intracellular and extracellular spaces, so it cannot by itself distinguish where the sodium is.

Separating compartments

Techniques such as inversion recovery can suppress signal from fluids with long T1, such as cerebrospinal fluid, so the remaining signal is weighted toward tissue sodium. Multi-quantum methods target the more restricted, bi-exponential sodium, but they are technically demanding and remain mainly research tools.

Limitations

  • Low spatial resolution, typically several millimetres per voxel.
  • Partial-volume effects blur small structures.
  • Quantification depends on calibration and corrections.
  • Sodium MRI is mainly a research method and not a routine clinical examination.

Key takeaways

  • Sodium reflects cellular energy and membrane function.
  • The signal is weak and decays quickly, so special readouts and long averaging are needed.
  • Total sodium concentration is a calibrated average, not a compartment-specific measure.
  • Ultra-high field improves signal-to-noise ratio for sodium MRI.

Frequently asked questions

Can sodium MRI replace proton MRI?

No. It complements it. Proton images provide anatomy and resolution, and sodium images add metabolic information at lower resolution.

Why is it done mainly at high field?

Signal-to-noise ratio increases with field strength, which helps a method that starts with very little signal.

References and further reading

  • Madelin G, Regatte RR. Biomedical applications of sodium MRI in vivo. J Magn Reson Imaging. 2013;38:511-529.
  • Madelin G, Lee JS, Regatte RR, Jerschow A. Sodium MRI: methods and applications. Prog Nucl Magn Reson Spectrosc. 2014;79:14-47.

Related articles

Last reviewed: 4 November 2026. Educational content; does not replace local protocols or manufacturer guidance.