ASL timing: post-labelling delay and arterial transit time

In arterial spin labelling, timing is everything. Labelled blood takes time to travel from the labelling plane to the tissue, and the label itself decays as it goes. Two timing settings, the labelling duration and the post-labelling delay (PLD), together with the physiological arterial transit time (ATT), decide whether the image shows tissue perfusion or something else. This article builds on Arterial spin labelling in plain English.

Learning objectives

  • Define labelling duration, post-labelling delay and arterial transit time.
  • Explain what happens when the PLD is too short or too long.
  • Describe why transit time varies between people and brain regions.
  • Recognise multi-delay ASL as a way to handle this variability.
Arterial spin labelling: labelled and control images and the resulting perfusion image
Arterial spin labelling: labelled and control images and the resulting perfusion image. Image: Alpheratzstar, Public domain, via Wikimedia Commons (File:Arterial_Spin_Labeling-experiment.png).

Three times to keep apart

TermWhat it isWho controls it
Labelling durationHow long blood is being labelledThe protocol
Post-labelling delay (PLD)Wait between end of labelling and image acquisitionThe protocol
Arterial transit time (ATT)Time for labelled blood to travel from the labelling plane to the tissueThe patient’s physiology
Timing diagram showing a labelling period, a post-labelling delay and then image acquisition
ASL timing: labelling duration, post-labelling delay and readout. Schematic only; real values depend on the protocol.

If the PLD is too short

When the PLD is shorter than the transit time, the labelled blood has not yet reached the tissue. It is still in the arteries. The image then shows bright vascular signal, and the tissue perfusion appears lower than it really is. This arterial transit artefact is common in older people and in patients with slow flow.

If the PLD is too long

The label decays with the longitudinal relaxation time of blood (T1), so every additional moment of delay costs signal. The consensus recommendations use a blood T1 of about 1.65 s at 3 T. A very long PLD therefore reduces signal-to-noise ratio in an already low-signal technique.

Typical recommendations

The ISMRM Perfusion Study Group and European ASL in Dementia consortium consensus for clinical use recommends PCASL with a labelling duration of about 1800 ms and a PLD of about 1800 ms for adults, and a somewhat longer PLD of about 2000 ms for older or clinical populations where transit times are longer (Alsop et al., 2015). Treat these as starting points and follow your local protocol.

Why transit time varies

  • Age: transit times tend to lengthen with age.
  • Disease: narrowed or occluded vessels and collateral flow delay arrival.
  • Brain region: territories further from the labelling plane, or supplied by slow collateral routes, receive blood later.
  • Physiological state: blood pressure, caffeine and other factors can change flow.

Multi-delay ASL

Instead of a single PLD, multi-delay ASL acquires images at several delays. Fitting a model across them gives an estimate of both cerebral blood flow and arterial transit time, at the cost of lower signal-to-noise ratio per delay or longer scans. Time-encoded approaches reduce this cost by encoding the delays within one acquisition.

Key takeaways

  • Labelling duration and PLD are protocol choices; transit time is physiological.
  • Too short a PLD leaves label in the arteries; too long a PLD loses signal.
  • Consensus values are starting points: around 1800 ms for healthy adults and longer for older or clinical populations.
  • Multi-delay ASL estimates transit time as well as perfusion.

Frequently asked questions

Why not just use a very long PLD for everyone?

The signal decays as the label relaxes, so a long PLD gives noisier images. The protocol is a trade-off.

Can ASL show regions of slow flow?

Yes, as delayed arrival, bright intravascular signal or reduced perfusion at short PLD. Interpretation should consider transit time before concluding that perfusion is reduced.

References and further reading

  • Alsop DC, et al. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications: a consensus of the ISMRM Perfusion Study Group and the European Consortium for ASL in Dementia. Magn Reson Med. 2015;73:102-116.

Related articles

Last reviewed: 4 November 2026. Educational content; does not replace local protocols or manufacturer guidance.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles

Back to top button