MRI Physics

Diffusion MRI Basics: b-values, DWI and the ADC Map

Learning objectives

  • Explain how diffusion gradients make MRI sensitive to water motion
  • Describe what the b-value controls
  • Calculate an apparent diffusion coefficient (ADC) from two images
  • Explain why DWI and ADC are read together
DTI tractography of a human brain, sagittal view
Diffusion data can be extended to tractography. DTI tractography of a human brain, sagittal view. Image: Thomas Schultz, CC BY-SA 3.0, via Wikimedia Commons (File:DTI-sagittal-fibers.jpg).

The idea in one paragraph

Water molecules move randomly. A diffusion-weighted sequence adds a pair of strong, balanced gradient pulses around the refocusing pulse (the Stejskal–Tanner design). Spins that stay still are fully rephased; spins that move between the two pulses are not, so they lose phase coherence and signal. The more freely water moves, the more signal is lost.

The b-value

The b-value (s/mm²) summarises how strongly the sequence weights diffusion. It grows with gradient strength, gradient duration and the time between the pulses. b = 0 gives a T2-weighted image with no diffusion weighting; clinical brain DWI commonly uses a high b-value around 1000 s/mm² (illustrative; protocols vary).

Line chart of relative signal versus b-value for three diffusion coefficients
Computed curves from the mono-exponential model S/S₀ = exp(−bD). D values are illustrative.

The signal equation

In the simplest (mono-exponential) model, signal at b relates to signal at b = 0 by:

S(b) = S0 · exp(−b · ADC)

Rearranging with two acquisitions gives the ADC for each voxel:

ADC = ln(S0 / Sb) / b

Worked example

Suppose a voxel has S0 = 1000 and S1000 = 450 at b = 1000 s/mm². Then ADC = ln(1000/450) / 1000 = ln(2.222)/1000 ≈ 0.799/1000 ≈ 0.80 × 10⁻³ mm²/s.

DWI versus ADC

ImageWhat bright meansWatch out for
DWI (high b)Less signal lost: more restricted diffusion, or long T2T2 shine-through: long-T2 tissue looks bright even without restriction
ADC mapFreer diffusion (higher ADC)Restricted diffusion appears dark

Because the DWI image still carries T2 weighting, a bright DWI area is confirmed as true restriction only when the ADC map is dark in the same place. See the pathology signal article for clinical context.

Why it matters

  • DWI is highly sensitive to early ischaemic change
  • Diffusion tensor imaging (DTI) extends the idea to direction-dependent diffusion in white matter
  • Diffusion models beyond mono-exponential (for example IVIM) separate perfusion-like effects at low b-values

Key takeaways

  • Diffusion gradients convert motion into signal loss
  • The b-value sets how strongly diffusion is weighted
  • ADC = ln(S0/Sb)/b; read DWI and ADC together

FAQ

Why is it called “apparent” diffusion coefficient?

Tissue is not free water: membranes, cells and perfusion all affect the measurement, so it is an apparent value rather than the true molecular diffusion coefficient.

Does a higher b-value always help?

Higher b increases diffusion contrast but lowers signal-to-noise ratio and can increase distortion, so b is a trade-off.

References

  • Stejskal EO, Tanner JE. Spin diffusion measurements: spin echoes in the presence of a time-dependent field gradient. J Chem Phys 1965;42:288-292.
  • Le Bihan D, Breton E, Lallemand D, et al. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology 1986;161:401-407.
  • Brown RW, Cheng YCN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design, 2nd ed. Wiley; 2014.

Educational disclaimer: values and parameters here are illustrative and vary by scanner, field strength, vendor and protocol. This article is for learning and is not clinical advice.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.