MRI Artefacts: Aliasing, Motion, Chemical Shift and Susceptibility
Learning objectives
- Recognise four common MRI artefacts
- Explain the physics behind each one
- Choose a practical fix and know its cost
Artefacts at a glance
| Artefact | Cause | Looks like | Common fixes |
|---|---|---|---|
| Aliasing (wrap-around) | Object larger than the field of view in the phase direction | Anatomy from outside the FOV folds onto the other side | Larger FOV, phase oversampling, swap phase direction, saturation bands |
| Motion and flow | Movement during the acquisition | Ghost copies along the phase-encoding direction | Immobilisation, faster sequences, gating, motion correction, flow compensation |
| Chemical shift | Fat and water precess at slightly different frequencies | Bright and dark bands at fat–water edges along the frequency direction | Higher bandwidth, fat suppression |
| Susceptibility | Field distortion near air, bone, metal or blood products | Signal loss and geometric distortion | Spin echo, shorter TE, higher bandwidth, metal-artefact techniques |
Aliasing
The phase-encoding steps sample k-space at a spacing set by the field of view. Anything outside that FOV is indistinguishable from signal inside it, so it folds back in. The examples below show what each artefact looks like on a scanner.




Examples of the four artefacts. Images: L.J. Erasmus et al., CC BY 4.0, via Wikimedia Commons.
See k-space explained for why the sample spacing sets the FOV.
Motion and flow
Each phase-encoding line is acquired at a different time. Motion between lines makes k-space inconsistent, which appears as ghosts repeated along the phase direction. Periodic motion such as breathing or pulsation gives regularly spaced ghosts.
Chemical shift
Fat resonates about 3.5 ppm below water. In hertz this grows with field strength, and the displacement in pixels depends on the receiver bandwidth per pixel:
shift (pixels) = fat–water difference (Hz) / bandwidth per pixel (Hz/pixel)
fat–water difference ≈ 3.5 ppm × Larmor frequency
| Bandwidth per pixel | 1.5 T (≈ 224 Hz) | 3 T (≈ 447 Hz) |
|---|---|---|
| 100 Hz/pixel | ≈ 2.2 pixels | ≈ 4.5 pixels |
| 200 Hz/pixel | ≈ 1.1 pixels | ≈ 2.2 pixels |
| 400 Hz/pixel | ≈ 0.6 pixels | ≈ 1.1 pixels |
Computed using 63.9 MHz at 1.5 T and 127.7 MHz at 3 T. Raising bandwidth reduces the shift but costs SNR (see SNR in MRI).
Susceptibility
Tissues and materials magnetise differently. At boundaries such as air–tissue near the sinuses, or around metal, the local field becomes non-uniform. Gradient-echo sequences are most sensitive because they do not refocus these static field offsets; SWI deliberately exploits this sensitivity to show blood products and iron.
Key takeaways
- Phase-direction artefacts (aliasing, ghosting) depend on how k-space is sampled
- Chemical shift scales with field strength and inversely with bandwidth
- Susceptibility is worst on gradient echo and near air or metal
FAQ
Why does swapping the phase and frequency directions help?
It moves aliasing and ghosts into a different direction, often away from the anatomy of interest, though it may change scan time or require oversampling.
Is every artefact a problem?
No. Some are useful: susceptibility effects underpin SWI, and chemical shift is used in Dixon fat–water imaging.
References
- Brown RW, Cheng YCN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design, 2nd ed. Wiley; 2014.
- Shepp LA, Logan BF. The Fourier reconstruction of a head section. IEEE Trans Nucl Sci 1974;21:21-43.
Educational disclaimer: values are illustrative and depend on scanner, field strength and protocol. Not clinical advice.