Denoising Sodium (²³Na) MRI: Averaging, Filtering, Non-Local Means and MP-PCA
Learning objectives
- Explain why sodium MRI is noise-limited and why magnitude noise is Rician
- Compare averaging, filtering and modern denoisers
- Run a basic denoising step in Python
- Judge the effect of denoising on tissue sodium concentration
Why sodium images are noisy
Tissue sodium is around a thousand times less concentrated than water protons and has lower MR sensitivity, so SNR is low even at 7 T. Voxels are therefore large and scans long. Noise is Gaussian in the real and imaginary channels; taking the magnitude makes it Rician, which adds a positive bias in low-signal regions and can inflate concentration estimates.

The options
| Method | Idea | Cost or risk |
|---|---|---|
| Averaging | SNR grows with √(number of averages) | Scan time; motion between averages |
| k-space filtering (Hamming, Hanning, Kaiser) | Down-weights noisy high frequencies | Blurs edges and increases partial volume |
| Non-local means (NLM) | Averages similar patches across the image | Can smooth small structures; set noise level carefully |
| BM4D | Collaborative filtering of grouped 3D blocks | Computationally heavier; parameter-sensitive |
| MP-PCA | Removes components below the Marchenko–Pastur noise threshold in redundant data (coils, echoes, repeats) | Needs enough redundant measurements |
| Regularised or compressed-sensing reconstruction | Denoises within the reconstruction | Can bias quantitative values if over-regularised |
| Deep learning | Learned mapping from noisy to clean | Risk of hallucinated detail; needs validation on sodium data |
Denoise before or after magnitude?
Where possible, denoise complex data (or each coil) before taking the magnitude. That avoids treating the Rician bias as signal. If only magnitude images are available, use methods with a Rician correction.
A minimal Python example
DIPY provides NLM and MP-PCA. For a 4D array (for example repeated sodium acquisitions stacked on the last axis):
import numpy as np
from dipy.denoise.localpca import mppca
from dipy.denoise.nlmeans import nlmeans
from dipy.denoise.noise_estimate import estimate_sigma
data = np.load("na_repeats.npy") # shape (x, y, z, n_repeats)
# MP-PCA uses the redundancy across repeats
den_mppca = mppca(data, patch_radius=2)
# NLM on the mean image, with Rician correction
mean_img = data.mean(axis=-1)
sigma = estimate_sigma(mean_img, N=1) # N = number of receive coils if known
den_nlm = nlmeans(mean_img, sigma=sigma, patch_radius=1, block_radius=2, rician=True)
Checking that denoising did no harm
- Scan reference phantoms of known concentration and process them identically to tissue
- Compare mean values in regions of interest before and after denoising; the mean should not shift beyond the noise level
- Inspect the residual (noisy − denoised): it should look like structure-free noise
- Report the method and parameters so results are reproducible
Gibbs ringing
Sodium images have few k-space samples, so truncation ringing near sharp edges (for example CSF boundaries) is common. Filtering reduces it at the cost of resolution; dedicated methods such as local subvoxel-shift unringing are an alternative.
Key takeaways
- Averaging is the safest way to gain SNR, but it costs time
- Every filter trades noise for resolution or bias; validate with phantoms
- MP-PCA needs redundant data; NLM and BM4D work on single images
FAQ
Which method should I start with?
A mild k-space filter applied identically to phantoms and tissue is the simplest reproducible baseline; add MP-PCA if you acquire repeats or multiple coils.
Does denoising change TSC?
It can. Smoothing mixes CSF (around 140 mmol/L) into nearby tissue, raising its apparent concentration, so always check regional means.
References
- Gudbjartsson H, Patz S. The Rician distribution of noisy MRI data. Magn Reson Med 1995;34:910-914.
- Buades A, Coll B, Morel JM. A non-local algorithm for image denoising. IEEE CVPR 2005;2:60-65.
- Manjón JV, Coupé P, Martí-Bonmatí L, Collins DL, Robles M. Adaptive non-local means denoising of MR images with spatially varying noise levels. J Magn Reson Imaging 2010;31:192-203.
- Maggioni M, Katkovnik V, Egiazarian K, Foi A. Nonlocal transform-domain filter for volumetric data denoising and reconstruction. IEEE Trans Image Process 2013;22:119-133.
- Veraart J, Novikov DS, Christiaens D, Ades-Aron B, Sijbers J, Fieremans E. Denoising of diffusion MRI using random matrix theory. NeuroImage 2016;142:394-406.
- Kellner E, Dhital B, Kiselev VG, Reisert M. Gibbs-ringing artifact removal based on local subvoxel-shifts. Magn Reson Med 2016;76:1574-1581.
- Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 2007;58:1182-1195.
Educational disclaimer: code and parameters are illustrative starting points. Validate any pipeline against phantoms and your own scanner data before using it for research results.