A protocol that looks fine on the scanner can still be a poor choice for the analysis you plan: a TE that is too long for BOLD at 3 T, a resting-state run that is too short, or voxels that do not match the field of view and matrix. The SpinLab Protocol Checker scores your acquisition parameters against published recommendations for your modality and field strength, checks the physics, and compares your protocol with large reference studies.

SpinLab Protocol Checker scoring a resting-state fMRI protocol
Protocol Checker scoring a 3T resting-state fMRI protocol at 100/100, with acquisition time, voxel volume, relative SNR efficiency and a relative SAR estimate. Detail from spinlab.mrispin.com.

Step by step

  1. Choose the study type: resting-state, task or multi-echo fMRI, structural T1 (MPRAGE), T2/FLAIR, diffusion (single or multi-shell), ASL, SWI, TOF angiography, spectroscopy or cardiac cine.
  2. Choose the field strength: 1.5 T, 3 T or 7 T.
  3. Enter your parameters, or start from a published protocol (HCP 3 T, HCP 7 T, a 7 T high-resolution protocol, UK Biobank or ABCD) and edit it.
  4. Read the score and the parameter checks. Each check shows your value, the recommended range and the reason.
  5. Open Physics & feasibility and Compare vs published.
  6. Export the protocol for Siemens, GE or Philips, generate a methods sentence, or save it.

Example: a 3 T resting-state protocol

With TR 800 ms, TE 30 ms, flip angle 52°, 2.4 mm isotropic voxels, 60 slices, multiband 6 and 750 volumes (10 minutes), all ten checks pass and the protocol scores 100/100. The physics panel reports:

MeasureValueHow it is calculated
Acquisition time10:00 minTR × volumes = 800 ms × 750
Voxel volume13.82 mm³2.4 × 2.4 × 2.4 mm
Relative SNR efficiency15.46Proportional to voxel volume ÷ √TR
Slice coverage144 mm60 slices × 2.4 mm
Excitation bands10 × 6Slices per band × multiband factor
Relative SAR estimate19% of typical limitProxy proportional to flip angle² × field² ÷ TR

The comparison table then sets your values beside HCP (TR 720 ms, 2 mm, MB 8), UK Biobank (TR 735 ms, 2.4 mm, MB 8) and ABCD (TR 800 ms, 2.4 mm, MB 6), with the percentage difference for each.

Example: what a problem looks like

SpinLab Protocol Checker flagging an echo time that is too long
With TE raised to 60 ms the score drops to 85 and the TE check fails, with the recommended 28–38 ms range and the reason. Detail from spinlab.mrispin.com.

Raising TE to 60 ms drops the score to 85 and the TE check fails. The explanation gives the reason: BOLD sensitivity is greatest when TE is close to the T2* of grey matter, about 30 ms at 3 T and about 22 ms at 7 T, so the recommended range here is 28–38 ms.

How to use the result

  • Use the checker when designing a study, before ethics approval, so that the protocol can be justified with references.
  • Treat the score as a guide, not a rule: a deliberate choice for your research question can be outside a typical range.
  • The SAR figure is a relative proxy for comparing protocols. The scanner’s own SAR monitoring always decides what can be run safely.

References

  • Glasser MF, Smith SM, Marcus DS, et al. The Human Connectome Project’s neuroimaging approach. Nat Neurosci 2016;19(9):1175–1187.
  • Miller KL, Alfaro-Almagro F, Bangerter NK, et al. Multimodal population brain imaging in the UK Biobank prospective epidemiological study. Nat Neurosci 2016;19(11):1523–1536.
  • Casey BJ, Cannonier T, Conley MI, et al. The Adolescent Brain Cognitive Development (ABCD) study: imaging acquisition across 21 sites. Dev Cogn Neurosci 2018;32:43–54.

Related articles

Last reviewed: 7 November 2026. SpinLab and SpinCalc are free tools developed by MRISpin. Screenshots show the tools as they were on this date and may change as the tools are updated. Generated scripts should be checked before use on real data.