MRI Liver and Upper Abdomen: Protocol, Positioning and Planning
This guide covers a routine liver and upper abdominal MRI from start to finish: anatomy and landmarks, indications, safety and preparation, positioning from coil to first image, and each sequence with its planning image and example parameters. For the bile ducts, see the dedicated MRCP guide.
Anatomy and planning landmarks
The liver lies under the right hemidiaphragm and extends across the midline to the left upper quadrant. It is divided into eight functional segments (Couinaud segments) by the three hepatic veins and the portal vein branches; lesions are reported by segment, so complete coverage and consistent planes matter.
For planning, the key landmarks are the dome of the diaphragm (upper limit), the lower edge of the right lobe and the lower poles of the kidneys (lower limit), and the position of the liver relative to the midline. Because the liver moves with breathing by several centimetres, every sequence is either acquired in a breath-hold or synchronised with breathing.
Indications
- Characterising a liver lesion found on ultrasound or CT
- Surveillance of patients with cirrhosis or chronic liver disease for hepatocellular carcinoma
- Detection and staging of liver metastases
- Assessment of fat and iron in the liver
- Assessment of the pancreas, spleen, adrenal glands and kidneys (with the appropriate protocol)
Safety screening
Every patient completes standard MRI safety screening before entering the scanner room, and any implant or device is checked against its MR Conditional labelling for your scanner. Ask specifically about surgical clips, biliary stents, transjugular shunts and embolisation coils. See MRI implant and device screening for the full process.
Patient preparation
- Fast for about 4 hours before the scan (follow local protocol); this reduces bowel movement and fills the gallbladder
- Practise breath-holds before the patient lies down, usually at end-expiration; consistent breath-holds are essential for dynamic imaging
- Screen renal function and allergies if gadolinium is planned, and check whether a liver-specific (hepatobiliary) agent such as gadoxetate is requested
- Place an IV cannula before positioning, in a position that allows the arms to be raised
- Remove all metal and change into a gown; record the patient’s weight
Positioning
| Step | Typical approach |
|---|---|
| Position | Supine, feet first |
| Coil | Anterior body array plus posterior table coil |
| Arms | Above the head if tolerated, to avoid wrap; otherwise by the sides on pads, away from the body |
| Centring | Lower costal margin / a few centimetres below the xiphoid |
| Breathing | Respiratory belt or navigator for free-breathing sequences |
Step by step: from coil to first image
- Check the request and screening: identity, indication, contrast agent, completed MRI safety screening and renal function.
- Prepare the table: posterior coil elements, sheet, knee support, and the respiratory belt if used.
- Lay the patient supine, feet first, and connect the IV line with enough slack for table movement.
- Place the arms above the head on a support if tolerated; otherwise by the sides on pads.
- Place the respiratory belt where the breathing movement is greatest, usually just below the ribs.
- Place the anterior body array over the upper abdomen, covering the liver and kidneys, and secure it with straps.
- Give the call bell and earplugs or headphones; repeat the breathing instructions.
- Switch on the laser and landmark at the lower costal margin, then move the table to isocentre.
- Register the patient on the console and select the coil elements.
- Run a breath-hold three-plane localiser and check that the whole liver is covered and inside the coil.




General photos of MRI equipment and workflow; your scanner, coils and room will look different. Images via Wikimedia Commons (credits in each image’s media details).
Liver MRI protocol and planning
The sequences below make up a typical examination. Sequence names differ between vendors (for example TSE/FSE, HASTE/SSFSE, SPACE/CUBE/VISTA). All parameters are illustrative examples, mostly for 1.5 T, and must be adapted to your scanner, coil and local protocol.
1. Three-plane localiser (breath-hold)
Planning: Acquire in a breath-hold. Check that the dome of the diaphragm and the lower edge of the liver are both covered and that the coil is centred over the liver.
2. Coronal T2 single-shot (HASTE / SSFSE), breath-hold

Planning: Plan on the axial localiser parallel to the left–right axis. Cover from the anterior abdominal wall to behind the kidneys, and check on the sagittal localiser that the stack covers the liver dome and lower edge.
| Parameter | Example value (illustrative) |
|---|---|
| Slice thickness / gap | 5–6 mm / 10% |
| TE | about 80–90 ms |
| Field of view | 380–400 mm |
| Breathing | One or two breath-holds |
3. Axial T1 in-phase and opposed-phase (Dixon), breath-hold

Planning: Plan on the coronal image, from the dome of the diaphragm to the lower poles of the kidneys (or the lower edge of the liver). Use the same coverage and position for all axial sequences.
Signal loss on opposed-phase images shows fat inside cells (for example hepatic steatosis); signal loss on the in-phase images (longer TE at 1.5 T) suggests iron.
| Parameter | Example value (illustrative) |
|---|---|
| Technique | 2D or 3D gradient echo, Dixon |
| TE (1.5 T) | about 2.2 ms (opposed) and 4.4 ms (in-phase) |
| Slice thickness | 5–6 mm (2D) or 3–4 mm (3D) |
| Breathing | Single breath-hold |
4. Axial T2 with fat suppression
Planning: Copy the axial geometry. Use a respiratory-triggered TSE or a breath-hold single-shot sequence depending on the patient.
| Parameter | Example value (illustrative) |
|---|---|
| Slice thickness / gap | 5–6 mm / 10% |
| TE | about 80–100 ms |
| Fat suppression | Spectral (SPAIR) or similar |
| Breathing | Respiratory-triggered or breath-hold |
5. Axial DWI with ADC map
Planning: Copy the axial geometry.
| Parameter | Example value (illustrative) |
|---|---|
| Technique | Single-shot EPI, fat-suppressed |
| b-values | for example 50, 400 and 800 s/mm² |
| Slice thickness | 5–6 mm |
| Breathing | Free-breathing, respiratory-triggered or navigator |
6. Axial 3D T1 fat-suppressed dynamic (pre- and post-contrast)
Planning: Copy the axial coverage. Acquire a pre-contrast image, then the arterial, portal venous and delayed phases after gadolinium, each in a breath-hold. Time the late arterial phase with bolus tracking or a test bolus.
| Parameter | Example value (illustrative) |
|---|---|
| Technique | 3D spoiled gradient echo with fat suppression (VIBE / LAVA / THRIVE) |
| Slice thickness | 2–3 mm (interpolated) |
| Phases (illustrative) | Pre-contrast, late arterial, portal venous (about 60–70 s), delayed (about 3–5 min) |
| Breathing | One breath-hold per phase |
7. Hepatobiliary phase (with gadoxetate only)
Planning: Repeat the axial 3D T1 about 20 minutes after injection of a hepatobiliary agent, with the same geometry; add a coronal if needed.
Tips and common pitfalls
- Inconsistent breath-holds misregister the dynamic phases; practise before the scan and give the same command every time.
- Time the arterial phase carefully; too early or too late reduces lesion conspicuity.
- Arms by the sides can wrap into the image; increase the field of view or use phase oversampling.
- Shim over the liver for uniform fat suppression.
- Keep the axial coverage identical for all sequences so that lesions can be compared slice by slice.
Frequently asked questions
Why do patients fast before liver MRI?
Fasting reduces bowel movement and fluid in the stomach and fills the gallbladder, which improves image quality.
What is the hepatobiliary phase?
With a hepatocyte-specific agent such as gadoxetate, normal liver cells take up the contrast; images at about 20 minutes help to characterise lesions.
References and further reading
- Westbrook C. Handbook of MRI Technique. 4th ed. Wiley-Blackwell; 2014.
- Westbrook C, Talbot J. MRI in Practice. 5th ed. Wiley-Blackwell; 2018.
- American College of Radiology. CT/MRI Liver Imaging Reporting and Data System (LI-RADS) v2018.
Related articles
Last reviewed: 6 November 2026. Educational content only. Positioning, coverage, sequences and parameters are illustrative examples. Always follow your department protocol, the referral question, MRI safety screening and the vendor recommendations for your scanner and coils.