Neuroimaging Software

Sodium MRI: Calculating Tissue Sodium Concentration with SpinLab

Sodium (²³Na) MRI measures something proton MRI cannot: the concentration of sodium ions in tissue. Healthy cells keep intracellular sodium low (around 10–15 mM) while the extracellular space sits near 140 mM, so a rise in tissue sodium concentration (TSC) is a sensitive marker of cell energy failure, oedema or a larger extracellular space. The catch is that sodium images are noisy and arrive in arbitrary signal units. The SpinLab Sodium TSC tool turns them into millimolar values using reference vials, and adds a simple compartment estimate on top.

Sodium tissue concentration calibration in SpinLab
Tissue sodium concentration from four reference vials, with a CSF correction and the extracellular volume fraction. From spinlab.mrispin.com.

Why calibration needs reference vials

The sodium signal depends on coil sensitivity, receiver gain, relaxation and the sequence, so a raw pixel value has no physical unit. The usual solution is to scan tubes of known sodium concentration (often saline in agar to mimic tissue relaxation) next to the head. Plotting the measured vial signal against the known concentration gives a straight line, and that line converts any tissue signal into millimolar.

  1. Place three or more vials with known sodium concentrations in the field of view.
  2. Draw a region of interest (ROI) in each vial and record the mean signal.
  3. Fit a straight line: signal = slope × concentration + intercept.
  4. Draw ROIs in tissue and invert the line: TSC = (signal − intercept) / slope.
  5. Check R². A good calibration is close to 1; a low value points to a misplaced ROI, a vial outside the coil’s even region, or partial volume.

Step by step in SpinLab

  1. Open spinlab.mrispin.com and choose Sodium TSC.
  2. Enter each vial’s known concentration and its measured mean signal.
  3. Enter the mean signal from your tissue ROI.
  4. Optionally set the tissue water fraction and the intra- and extracellular sodium values for the compartment estimate.
  5. Read the fit (slope, intercept, R²), the TSC in mM and the estimated extracellular volume fraction.

In the example shown above, the calibration fit gave a slope of 2.008, an intercept of 0.90 and R² = 0.9995, and the tissue ROI came out at 41.9 mM.

From TSC to extracellular volume

A common two-compartment model treats tissue sodium as the sum of a small intracellular pool and a large extracellular pool, scaled by the water fraction w:

TSC = w × [ ECV × Na_e + (1 − ECV) × Na_i ]

ECV = (TSC / w − Na_i) / (Na_e − Na_i)

With w = 0.8, Nae = 140 mM and Nai = 12 mM, a TSC of 41.9 mM gives an extracellular volume fraction (ECV) of about 32%. Treat this as an estimate: it depends on the assumed water fraction and concentrations, and it ignores relaxation differences between the two pools.

Do the same in Python

The short script below repeats the calculation. The vial and ROI numbers are illustrative values chosen for this example, not data from a real scan.

import numpy as np

# Illustrative values: known vial concentrations (mM) and mean ROI signal
conc   = np.array([20, 40, 60, 80])
signal = np.array([41.0, 81.2, 121.4, 161.7])

slope, intercept = np.polyfit(conc, signal, 1)
r2 = np.corrcoef(conc, signal)[0, 1] ** 2

tissue_signal = 85.0
tsc = (tissue_signal - intercept) / slope

w, na_e, na_i = 0.8, 140, 12
ecv = (tsc / w - na_i) / (na_e - na_i)

print(f"slope={slope:.3f}  intercept={intercept:.2f}  R2={r2:.4f}")
print(f"TSC = {tsc:.1f} mM   ECV = {ecv:.0%}")
slope=2.011  intercept=0.75  R2=1.0000
TSC = 41.9 mM   ECV = 32%

Practical tips

  • Keep the vials inside the region where the coil is most uniform, or correct for coil sensitivity before fitting.
  • Use concentrations that bracket the tissue values you expect (brain tissue is roughly 30–50 mM; CSF is near 140 mM).
  • Sodium has very short T2* components. Ultra-short echo time sequences reduce the signal you lose before readout.
  • Small ROIs in sodium images suffer from partial volume with CSF. The tool’s CSF correction helps, but careful ROI placement helps more.

More SpinLab guides: ASL CBF quantification and the SpinLab overview.

Last reviewed: 7 November 2026. SpinLab and SpinCalc are free tools developed by MRISpin. Screenshots show the tools as they were on this date and may change as the tools are updated. Generated scripts should be checked before use on real data.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles