BOLD fMRI explained: from blood oxygenation to activation maps
Learning objectives
- Explain where the BOLD signal comes from
- Describe the haemodynamic response and its timing
- Choose sensible echo time and repetition time for BOLD fMRI
- Recognise the main artefacts and confounds in fMRI

Blood as a natural contrast agent
Haemoglobin changes its magnetic properties with oxygen. Oxyhaemoglobin is diamagnetic, close to the surrounding tissue, while deoxyhaemoglobin is paramagnetic. Deoxyhaemoglobin creates tiny field disturbances around vessels, which speed up T2* decay and lower the gradient-echo signal. Ogawa and colleagues described this blood oxygenation level dependent (BOLD) contrast in 1990.
Why active brain becomes brighter
When neurons become active, local blood flow increases more than oxygen use. The extra oxygenated blood washes out deoxyhaemoglobin, T2* lengthens slightly, and the signal rises. The change is small: typically a few per cent at 3 T, and larger at 7 T.
The haemodynamic response
The BOLD response is slow and delayed. After a brief event it peaks after about 5–6 seconds and is often followed by a small undershoot before returning to baseline. Analysis software models this with a haemodynamic response function (HRF). A common choice is the double-gamma function:
import numpy as np
from scipy.stats import gamma
t = np.arange(0, 32, 0.1) # seconds
hrf = gamma.pdf(t, 6) - gamma.pdf(t, 16) / 6
hrf /= hrf.max()
print(t[hrf.argmax()]) # peak at 5.0 s
print(t[hrf.argmin()]) # undershoot minimum at about 15.7 s
In a GLM analysis, each condition’s timing is convolved with this function to predict the signal, and the fit tells you where the brain followed the task. See FSL design files explained for how this is set up in FSL.

Acquisition: gradient-echo EPI
BOLD fMRI almost always uses single-shot gradient-echo echo-planar imaging (EPI), which acquires a whole slice in tens of milliseconds. The echo time is set close to the T2* of grey matter to maximise BOLD sensitivity.
| Parameter | Typical value | Note |
|---|---|---|
| TE | ≈ 30 ms at 3 T (longer at 1.5 T, shorter at 7 T) | Close to grey matter T2* |
| TR | ≈ 0.7–2 s | Shorter with simultaneous multi-slice (multiband) |
| Voxel size | ≈ 2–3 mm isotropic | Smaller at 7 T |
| Volumes | Hundreds per run | Depends on the task design |
Artefacts and confounds
- Distortion and signal loss near the sinuses and ear canals (orbitofrontal cortex, inferior temporal lobes). Field maps or reversed phase-encoding scans are used for correction.
- Head motion, which can create false activation, especially if it is linked to the task.
- Physiological noise from breathing and heartbeat.
- Vascular origin. BOLD measures a blood response, not neurons directly, and large draining veins can displace activation.
References
- Ogawa S, Lee TM, Kay AR, Tank DW. Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc Natl Acad Sci USA 1990;87(24):9868–9872.
- Logothetis NK. What we can do and what we cannot do with fMRI. Nature 2008;453:869–878.
Related articles
Last reviewed: 7 November 2026. Educational content for researchers, clinicians and students. Values are typical examples; local protocols, scanner software and manufacturer guidance take priority.