Fat suppression in MRI: STIR, fat saturation, SPAIR and Dixon compared
Learning objectives
- Explain why fat is bright on many MRI sequences and why we suppress it
- Compare chemical fat saturation, STIR, SPAIR and Dixon
- Choose the right method near metal, after contrast and in large fields of view
- Combine Dixon in-phase and opposed-phase images into water and fat images

Why suppress fat?
Fat has a short T1 and a fairly long T2, so it is bright on T1-weighted images and on fast spin-echo T2-weighted images. That brightness can hide what we are looking for: bone marrow oedema, fluid in a joint, an enhancing lesion after gadolinium, or inflammation in soft tissue. Fat suppression darkens the fat so that water-based abnormalities stand out.
All fat suppression methods use one of two differences between fat and water: their resonance frequency (chemical shift) or their T1.
The chemical shift between fat and water
The main fat peak resonates about 3.5 parts per million (ppm) below water. In hertz, this difference grows with the field strength:
| Field strength | Larmor frequency | Fat–water difference (≈3.5 ppm) |
|---|---|---|
| 1.5 T | 63.9 MHz | ≈ 224 Hz |
| 3 T | 127.7 MHz | ≈ 447 Hz |
| 7 T | 298.0 MHz | ≈ 1,043 Hz |
A bigger difference makes frequency-selective fat saturation easier at 3 T than at 1.5 T, and harder on low-field systems.
1. Chemical (spectral) fat saturation
A frequency-selective RF pulse tuned to fat tips the fat magnetisation into the transverse plane, and a spoiler gradient destroys it just before the imaging pulse. Water is left untouched. This is often called CHESS or simply “fat sat”.
- Strengths: specific to fat, so it can be used after gadolinium; keeps good SNR.
- Weaknesses: depends on a uniform B0. Near metal, at the edges of a large field of view, at the neck and shoulder, or where air meets tissue, the fat frequency shifts and suppression fails, often patchily. Poor shimming can even suppress water by mistake.
2. STIR (short tau inversion recovery)
STIR uses T1 rather than frequency. A 180° inversion pulse is followed by the excitation pulse at the moment fat magnetisation passes through zero, the null point. For a long TR, the null point is TI ≈ 0.69 × T1 of fat, which in practice is about 150–180 ms at 1.5 T and roughly 200–260 ms at 3 T.

- Strengths: works even with poor B0 homogeneity, near metal and over large fields of view, so it is the workhorse for spine, marrow and whole-body imaging.
- Weaknesses: lower SNR, and it is not fat-specific: anything with a T1 close to fat is also nulled. Tissue that has taken up gadolinium has a shortened T1, so STIR should not be used after contrast.
3. SPAIR and SPIR (hybrid methods)
Spectral attenuated inversion recovery (SPAIR; vendors use their own names for similar methods) uses a frequency-selective, adiabatic inversion pulse applied only to fat, with the TI set to fat’s null point. It is fat-specific like chemical saturation but less sensitive to B1 variations, and is popular for breast and abdominal imaging at 3 T. It still depends on good B0 homogeneity.
4. Dixon (water–fat separation)
Dixon methods acquire images at echo times where fat and water are in phase and out of phase (opposed phase), then calculate separate water and fat images. One acquisition gives four image types: in-phase, opposed-phase, water-only and fat-only.
| Field strength | Opposed phase (first) | In phase (first) |
|---|---|---|
| 1.5 T | ≈ 2.3 ms | ≈ 4.6 ms |
| 3 T | ≈ 1.15 ms | ≈ 2.3 ms |

In the simplest two-point version, with water signal W and fat signal F:
In-phase (IP) = W + F
Opposed (OP) = W − F
Water = (IP + OP) / 2
Fat = (IP − OP) / 2
Fat fraction = Fat / (Water + Fat)
A quick check in Python, using illustrative signal values for a voxel containing both water and fat:
ip, op = 140.0, 60.0 # illustrative in-phase and opposed-phase signals
water = (ip + op) / 2 # 100.0
fat = (ip - op) / 2 # 40.0
ff = fat / (water + fat) # 0.286
print(water, fat, round(ff * 100, 1), "%")
Real scanners use more echoes and correct for B0 inhomogeneity and T2* decay, which makes Dixon robust over large fields of view and allows fat fraction mapping of the liver, muscle and marrow. A known pitfall is the fat–water swap, where the algorithm labels a region the wrong way round, usually at the edge of the field of view.
Choosing a method
| Situation | Usually best | Why |
|---|---|---|
| After gadolinium | Chemical fat sat, SPAIR or Dixon | STIR would also null enhancing tissue |
| Near metal implants | STIR (or Dixon) | Independent of fat frequency |
| Large field of view, spine, whole body | STIR or Dixon | Tolerant of B0 inhomogeneity |
| Small joints, high resolution | Chemical fat sat | Higher SNR, good shim over a small area |
| Fat quantification, adrenal adenoma, steatosis | Dixon (in and opposed phase) | Measures signal loss and fat fraction |
Key points
- Fat suppression uses either chemical shift (fat sat, SPAIR, Dixon) or T1 (STIR).
- Frequency-selective methods fail where B0 is not uniform.
- STIR is robust but non-specific; avoid it after contrast.
- Dixon gives water, fat, in-phase and opposed-phase images from one scan.
References
- Delfaut EM, Beltran J, Johnson G, Rousseau J, Marchandise X, Cotten A. Fat suppression in MR imaging: techniques and pitfalls. RadioGraphics 1999;19(2):373–382.
- Bley TA, Wieben O, François CJ, Brittain JH, Reeder SB. Fat and water magnetic resonance imaging. J Magn Reson Imaging 2010;31(1):4–18.
Related articles
- T1 and T2 relaxation explained
- MRI artefacts: aliasing, motion, chemical shift and susceptibility
- Larmor frequency explained
Last reviewed: 7 November 2026. Educational content for researchers, clinicians and students. Values are typical examples; local protocols, scanner software and manufacturer guidance take priority.