T1 and T2 relaxation explained: why tissues look different on MRI

Two tissues placed in the same scanner can give very different signals. The reason is relaxation: the way the magnetisation returns to equilibrium after an RF pulse. T1 and T2 are the time constants that describe it, and most conventional MRI contrast is built from them.

Learning objectives

  • Define T1 (longitudinal) and T2 (transverse) relaxation.
  • Explain how repetition time (TR) and echo time (TE) turn those differences into image contrast.
  • Recognise T1-weighted, T2-weighted and proton-density-weighted images by what is bright and dark.
  • Distinguish T2 from T2*.

What happens after an RF pulse

As covered in Larmor frequency explained, an RF pulse at the Larmor frequency tips the net magnetisation away from the static field B₀. Afterwards two things happen at once. The component along B₀ recovers (T1 relaxation), and the component perpendicular to B₀ fades away (T2 relaxation).

Two line charts: T1 recovery curves rising towards one and T2 decay curves falling towards zero for two example tissues with different time constants
Illustrative T1 recovery and T2 decay curves. Schematic with approximate values, not measured data.

T1: longitudinal recovery

T1 is the time constant for the longitudinal magnetisation to regrow toward its equilibrium value. After a pulse that fully tips the magnetisation, the recovery follows 1 − e−t/T1. Tissues with a short T1 recover quickly; tissues with a long T1 recover slowly. Fluids such as cerebrospinal fluid have long T1 values, while fat has a short one.

Mz(t) = M0 × (1 − e^(−t / T1))

T2: transverse decay

T2 is the time constant for the transverse magnetisation to decay. It reflects interactions between neighbouring spins that make them lose phase coherence. After a pulse, the transverse signal follows e−t/T2. Fluids have long T2 values and appear bright on T2-weighted images.

Mxy(t) = M0 × e^(−t / T2)

T2 versus T2*

In practice the signal decays faster than T2 predicts, because the static field is never perfectly uniform and because of susceptibility differences between tissues. The combined decay is described by T2*, which is always shorter than or equal to T2. Spin-echo sequences refocus the static-field part of the dephasing and are governed mainly by T2; gradient-echo sequences are sensitive to T2*. See spin echo versus gradient echo.

Turning relaxation into contrast

Two timing parameters let the operator choose which property dominates. TR is the time between successive excitations; TE is the time from excitation to the echo that is read out.

WeightingTRTEWhat it emphasises
T1-weightedShortShortDifferences in T1 recovery
T2-weightedLongLongDifferences in T2 decay
Proton-density-weightedLongShortDifferences in the number of mobile hydrogen nuclei

“Short” and “long” are relative to the tissue time constants, and actual values depend on field strength, sequence and vendor. They are not fixed numbers.

How the field strength matters

Relaxation times depend on field strength and tissue. In general, T1 values lengthen as the field increases, while T2 values change much less. Published measurements at 3 T place gray matter T1 at roughly 1.3 s and white matter roughly 0.8 s, with T2 values of roughly 110 ms and 80 ms. These are approximate, vary between studies, and are used in our figure only as illustrative numbers.

Key takeaways

  • T1 describes recovery along B₀; T2 describes loss of phase coherence in the transverse plane.
  • TR and TE are the main controls for weighting.
  • T2* includes field inhomogeneity and is shorter than T2.
  • Quoted tissue values are approximate and depend on field strength.

Frequently asked questions

Why is CSF dark on T1 and bright on T2 images?

CSF has long T1 and long T2. Its slow T1 recovery gives low signal when TR is short, while its slow T2 decay leaves high signal at long TE.

Are the curves in the figure real data?

No. They are drawn from the equations above with approximate time constants, to illustrate the shape of the curves.

References and further reading

  • Bloch F. Nuclear induction. Phys Rev. 1946;70:460-474.
  • Brown RW, Cheng YCN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design. 2nd ed. Wiley; 2014.
  • Bojorquez JZ, et al. What are normal relaxation times of tissues at 3 T? Magn Reson Imaging. 2017;35:69-80.

Related articles

Last reviewed: 4 November 2026. Educational content; does not replace local protocols or manufacturer guidance.