ASL CBF quantification with SpinLab: the white-paper model, Python code and oxford_asl

Arterial spin labelling (ASL) measures cerebral blood flow (CBF) without contrast, but turning the images into numbers needs a model and several constants. The ASL & Sodium Toolkit in SpinLab does this for you: it quantifies CBF from a single post-labelling delay (PLD) with the white-paper model, fits CBF and arterial transit time from multi-delay data, plans PLDs and background-suppression timing, writes BIDS ASL sidecars and gives the matching FSL oxford_asl command.

SpinLab ASL CBF calculator using the white-paper single-delay model
Single-delay CBF quantification for pCASL at 3 T using the ASL white-paper model, with the matching oxford_asl (FSL BASIL) command. Detail from spinlab.mrispin.com.

Step by step: single-delay CBF

  1. Open the toolkit and choose CBF (single delay).
  2. Choose the labelling scheme (pCASL, CASL or PASL) and the field strength. The field strength sets the white-paper blood T1 (1.65 s at 3 T).
  3. Enter the mean ΔM/M0 (control minus label, divided by M0) in percent, the PLD and the labelling duration τ.
  4. Check the constants: labelling efficiency α (0.85 for pCASL; lower it if background-suppression pulses are used), partition coefficient λ (0.9 mL/g) and, optionally, haematocrit to set blood T1.
  5. Enter the M0 repetition time and tissue T1 so that the toolkit can correct an M0 acquired with a short TR.
  6. Read the estimated CBF, compare it with the typical grey- and white-matter ranges, and copy the oxford_asl command and methods text.

The model

For pCASL and CASL, the consensus white paper (Alsop et al., 2015) uses a single-compartment model:

CBF = 6000 · λ · (ΔM/M0) · e^(PLD/T1b)
      ─────────────────────────────────────
      2 · α · T1b · (1 − e^(−τ/T1b))
SymbolMeaningTypical value
λBlood–brain partition coefficient0.9 mL/g
ΔM/M0Perfusion-weighted signal relative to the equilibrium magnetisationabout 0.5–1% in grey matter
PLDPost-labelling delay1.8 s (adults, pCASL)
τLabelling duration1.8 s
T1bT1 of arterial blood1.65 s at 3 T
αLabelling efficiency0.85 (pCASL)
6000Converts mL/g/s to mL/100 g/min—

The term ePLD/T1b corrects for the decay of the label while it waits for the delay, and the denominator accounts for the labelling duration and efficiency. The result is in mL/100 g/min.

Check the numbers yourself in Python

import numpy as np

def cbf_pcasl(dm_m0, pld=1.8, tau=1.8, t1b=1.65, alpha=0.85, lam=0.9):
    """White-paper single-PLD CBF (mL/100 g/min). dm_m0 as a fraction."""
    num = 6000 * lam * dm_m0 * np.exp(pld / t1b)
    den = 2 * alpha * t1b * (1 - np.exp(-tau / t1b))
    return num / den

def m0_tr_factor(tr=6.0, t1_tissue=1.5):
    """Factor to correct an M0 acquired with a finite TR."""
    return 1 / (1 - np.exp(-tr / t1_tissue))

cbf = cbf_pcasl(0.007)            # ΔM/M0 = 0.7 %
print(round(cbf, 1))               # 60.4
print(round(cbf / m0_tr_factor(), 1))  # 59.3, as shown in SpinLab

With ΔM/M0 = 0.7%, PLD = τ = 1.8 s at 3 T, the model gives 60.4 mL/100 g/min. The M0 image was acquired with TR = 6 s, so it is about 1.9% below full equilibrium (factor 1.019); correcting M0 lowers CBF to 59.3, the value in the screenshot, inside the typical grey-matter range of about 50–70.

The oxford_asl command, explained

oxford_asl -i asl.nii.gz -o oxasl_out \
  --iaf=tc --ibf=rpt \
  --casl --bolus=1.8 --plds=1.8 \
  -c M0.nii.gz --tr=6 --cmethod=voxel \
  --alpha=0.85 --t1b=1.65 \
  --spatial --mc
OptionMeaning
--iaf=tcVolumes are label–control pairs, tag first
--ibf=rptVolumes are grouped by repeat
--caslContinuous or pseudo-continuous labelling
--bolus, --pldsLabelling duration and PLD in seconds
-c, --trThe M0 calibration image and its TR
--cmethod=voxelCalibrate voxel by voxel with M0
--alpha, --t1bLabelling efficiency and blood T1, matching the values above
--spatialAdaptive spatial regularisation of the perfusion map
--mcMotion correction

Add --fslanat=<fsl_anat output> for registration to the structural image and partial-volume correction.

Multi-delay ASL

SpinLab multi-delay ASL fit of CBF and arrival time
Multi-delay ASL: fitting CBF and arterial transit time from several post-labelling delays. Detail from spinlab.mrispin.com.

With several PLDs, the Multi-delay fit tab fits CBF and arterial transit time (ATT) together, using the general kinetic model (Buxton et al., 1998). In the example data with seven delays from 0.25 to 2.5 s, it fits 60.2 ± 0.4 mL/100 g/min and an ATT of 1210 ms, with R² = 0.995. Measuring ATT matters in older people and in vascular disease, where a single PLD can be too short for the label to arrive.

Tips

  • Use the same constants in SpinLab, in oxford_asl and in your methods section.
  • Lower α when background-suppression pulses are used, because each pulse slightly reduces labelling efficiency.
  • Typical grey-matter and white-matter ranges are guides, not diagnostic thresholds.

References

  • Alsop DC, Detre JA, Golay X, et al. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications. Magn Reson Med 2015;73(1):102–116.
  • Buxton RB, Frank LR, Wong EC, et al. A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med 1998;40(3):383–396.
  • Chappell MA, Groves AR, Whitcher B, Woolrich MW. Variational Bayesian inference for a nonlinear forward model. IEEE Trans Signal Process 2009;57(1):223–236.

Related articles

Last reviewed: 7 November 2026. SpinLab and SpinCalc are free tools developed by MRISpin. Screenshots show the tools as they were on this date and may change as the tools are updated. Generated scripts should be checked before use on real data.