Susceptibility-weighted imaging (SWI): how it works and when to use it
Learning objectives
- Explain what magnetic susceptibility is and why blood products and veins appear dark
- Describe how SWI combines magnitude and phase images
- Know the main clinical uses of SWI
- Recognise blooming and the limits of phase interpretation

Susceptibility in one paragraph
Every material becomes slightly magnetised in the scanner. Paramagnetic substances, such as deoxyhaemoglobin, haemosiderin and ferritin, strengthen the local field. Diamagnetic substances, such as calcium and most tissue, weaken it slightly. These tiny differences distort the field around them, so nearby spins lose phase coherence and the signal on gradient-echo images drops. Susceptibility-weighted imaging (SWI) is designed to make the most of this effect.
How SWI is made
- A 3D gradient-echo sequence with a long TE (for example about 20 ms at 3 T and about 40 ms at 1.5 T), usually flow-compensated so moving blood does not cause artefacts.
- Magnitude image: regions with field distortion lose signal (T2* effect).
- Phase image: shows the local field shift directly. It is high-pass filtered to remove slow background changes.
- Phase mask: the filtered phase is turned into a mask that darkens pixels with paramagnetic phase, and the magnitude image is multiplied by this mask several times.
- Minimum intensity projection (minIP): thin slabs are projected so that veins can be followed as continuous dark lines.
Vendors use their own names for similar methods, such as SWI, SWAN and SWIp.
What SWI is good at
| Use | Why SWI helps |
|---|---|
| Microbleeds | Small haemosiderin deposits appear as round dark dots, much more visible than on standard T2* |
| Cavernomas | Blood products give strong signal loss, and multiple small lesions are easier to count |
| Traumatic brain injury | Small haemorrhages from diffuse axonal injury |
| Veins and venous thrombosis | Deoxygenated blood in veins is dark; a thrombosed vessel or slow flow can be seen |
| Acute stroke | The clot can appear as a dark “susceptibility vessel sign”, and prominent veins may suggest tissue at risk |
| Calcium versus blood | Phase can separate diamagnetic calcium from paramagnetic blood products |
| Multiple sclerosis | A vein running through the centre of a lesion (central vein sign) |
| Anti-amyloid therapy monitoring | Microbleeds and superficial siderosis (ARIA-H) |

Blooming and other pitfalls
- Blooming: signal loss extends beyond the true edge of a lesion, so size is overestimated. Blooming increases with field strength and TE, and it is also seen on other gradient-echo and echo-planar sequences, as in the image above.
- Phase sign depends on the scanner: whether paramagnetic material appears bright or dark on phase depends on the vendor’s conventions. Check with a known vein before calling calcium or blood.
- Air and bone: large susceptibility differences near the sinuses and skull base cause signal loss that can hide lesions.
- Mimics: flow voids, calcification and air can all be dark.
Quantitative susceptibility mapping (QSM) goes a step further by calculating the actual susceptibility from the phase, which removes the blooming and the vendor sign problem.
Key points
- SWI uses a long-TE 3D gradient echo and combines magnitude with a filtered phase mask.
- It is very sensitive to blood products, veins and iron.
- Expect blooming and check the phase convention before interpreting calcium versus blood.