Body Imaging PlanningMRI Planning

MRI Pelvis and Prostate: Protocol, Positioning and Planning

This guide covers pelvic MRI with a focus on the prostate, from start to finish: anatomy and landmarks, indications, safety and preparation, positioning from coil to first image, and each sequence with its planning image and example parameters. Notes on the female pelvis are included at the end.

Anatomy and planning landmarks

The prostate lies below the bladder and in front of the rectum, with its base against the bladder neck and its apex pointing down towards the pelvic floor. On T2-weighted images the peripheral zone is normally bright, and the transition zone around the urethra is darker and more heterogeneous; most cancers arise in the peripheral zone. The seminal vesicles sit above and behind the prostate.

For planning, the key landmarks are the symphysis pubis (the usual centring point), the bladder neck, the rectum and the long axis of the prostate on the sagittal image. In the female pelvis, the uterus, cervix and ovaries are planned in relation to the long and short axes of the uterus or cervix.

Indications

  • Detection and localisation of prostate cancer, including before biopsy
  • Staging of known prostate cancer and active surveillance
  • Rectal cancer staging (dedicated protocol)
  • Female pelvis: endometrial and cervical cancer staging, fibroids, adenomyosis, endometriosis, adnexal masses
  • Pelvic pain, fistula and congenital anomalies (dedicated protocols)

Safety screening

Every patient completes standard MRI safety screening before entering the scanner room, and any implant or device is checked against its MR Conditional labelling for your scanner. Ask specifically about hip replacements, which can cause artefact, and about penile or urinary implants. See MRI implant and device screening for the full process.

Patient preparation

  • Ask the patient to empty the rectum before the scan; some departments use a small enema or a short rectal preparation to reduce gas and stool
  • An antispasmodic (for example hyoscine butylbromide or glucagon) may be given to reduce bowel movement if not contraindicated, according to local protocol
  • PI-RADS v2.1 notes that avoiding ejaculation for about three days beforehand may help distension of the seminal vesicles; follow local advice
  • A moderately full bladder is usually preferred; ask the patient to void about an hour before the scan rather than immediately before
  • Screen renal function and allergies if contrast (DCE) is planned and place an IV cannula
  • Remove all metal and change into a gown; record the patient’s weight

Positioning

StepTypical approach
PositionSupine, feet first
CoilAnterior pelvic or body array plus posterior table coil (endorectal coil at some sites)
CentringSymphysis pubis
ComfortPad under the knees; arms on the chest or above the head, away from the pelvis

Step by step: from coil to first image

  1. Check the request and screening: identity, indication, contrast needs, completed MRI safety screening and renal function.
  2. Check preparation: rectal evacuation and timing of bladder emptying.
  3. Prepare the table: posterior coil elements, sheet and knee support.
  4. Lay the patient supine, feet first, and connect the IV line if contrast is planned.
  5. Place the anterior pelvic array over the pelvis so that its centre is at the symphysis pubis, and secure it with straps.
  6. Place the arms on the chest or above the head, clear of the coil and the pelvis.
  7. Give the call bell and earplugs or headphones; give the antispasmodic if used, just before scanning.
  8. Switch on the laser and landmark at the symphysis pubis, then move the table to isocentre.
  9. Register the patient on the console and select the coil elements.
  10. Run the three-plane localiser and check that the prostate (or uterus) is at the centre of the coil.

General photos of MRI equipment and workflow; your scanner, coils and room will look different. Images via Wikimedia Commons (credits in each image’s media details).

Pelvis and prostate MRI protocol and planning

The sequences below make up a typical examination. Sequence names differ between vendors (for example TSE/FSE, HASTE/SSFSE, SPACE/CUBE/VISTA). All parameters are illustrative examples, mostly for 1.5 T, and must be adapted to your scanner, coil and local protocol.

1. Three-plane localiser

Planning: Check that the prostate is at the centre of the coil and that the whole pelvis is in the field of view.

2. Large-field axial T1 (and T2) of the pelvis

Pelvis MRI large-field axial planning on a coronal image
Large-field axial slices planned on a coronal image, from the iliac crests to below the symphysis pubis. Coronal T1: Nevit Dilmen, CC BY-SA 3.0 (T1 channel shown in grey scale); axial T2: Nevit Dilmen, CC BY-SA 3.0 (T2 channel shown in grey scale); images from two different patients, via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Plan on the coronal localiser, from the aortic bifurcation or iliac crests to below the symphysis pubis. This series shows lymph nodes and bones for staging.

ParameterExample value (illustrative)
Slice thickness / gap5 mm / 10%
TR / TE (T1)about 500–700 ms / 10–15 ms
Field of view350–400 mm

3. Small-FOV axial T2 of the prostate

Prostate MRI small field-of-view axial planning
Small field-of-view axial block placed over the prostate on the coronal image, with the field of view centred on the prostate on the axial image. Coronal T1: Nevit Dilmen, CC BY-SA 3.0 (T1 channel shown in grey scale); axial T2: Nevit Dilmen, CC BY-SA 3.0 (T2 channel shown in grey scale); images from two different patients, via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Plan on the sagittal T2, either straight axial or perpendicular to the long axis of the prostate (or to the rectal wall), covering from below the apex to above the seminal vesicles. Check on the coronal image that the block is straight. Keep the same angulation and position for DWI and DCE so that all series match.

ParameterExample value (illustrative)
Slice thickness / gap3 mm / 0
TR / TEabout 3500–5000 ms / 100–120 ms
Field of view12–20 cm
In-plane resolution≤0.7 mm (phase) × ≤0.4 mm (frequency), as recommended by PI-RADS v2.1

4. Sagittal and coronal T2 of the prostate

Prostate MRI sagittal and coronal planning on an axial image
Sagittal and coronal T2 blocks planned across the prostate on an axial T2 image. Coronal T1: Nevit Dilmen, CC BY-SA 3.0 (T1 channel shown in grey scale); axial T2: Nevit Dilmen, CC BY-SA 3.0 (T2 channel shown in grey scale); images from two different patients, via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Sagittal: plan on the axial image parallel to the midline, covering the whole gland and seminal vesicles. Coronal: plan parallel to the long axis of the prostate on the sagittal image, covering the gland from front to back.

ParameterExample value (illustrative)
Slice thickness / gap3 mm / 0
Field of view12–20 cm

5. Axial DWI with ADC map (and high b-value images)

Planning: Copy the small-FOV axial T2 geometry.

ParameterExample value (illustrative)
TechniqueSingle-shot EPI, fat-suppressed
b-values (ADC)for example 50–100 and 800–1000 s/mm²
High b-value≥1400 s/mm² (acquired or calculated)
Slice thickness≤4 mm, no gap

6. Axial dynamic contrast-enhanced (DCE) T1

Planning: Copy the axial geometry. Start a series of fast 3D T1 acquisitions before injection and continue after gadolinium.

ParameterExample value (illustrative)
Technique3D spoiled gradient echo, fat-suppressed
Temporal resolution≤15 s per phase (PI-RADS v2.1)
Total durationat least 2 minutes
Slice thickness3 mm

Female pelvis notes

  • Plan sagittal T2 first through the midline of the uterus.
  • For the uterus, plan oblique axial (short axis) and oblique coronal (long axis) T2 images relative to the endometrial cavity on the sagittal image.
  • For the cervix, plan oblique axial images perpendicular to the long axis of the cervix.
  • A moderately filled bladder and an antispasmodic improve image quality; follow the dedicated gynaecological protocol.

Tips and common pitfalls

  • Rectal gas distorts DWI; preparation and correct positioning help.
  • Keep the field of view small for T2, but include at least one large-field axial series to assess lymph nodes and bones.
  • Patient movement between series misaligns T2, DWI and DCE; make the patient comfortable before starting.
  • Hip replacements cause artefact; adjust the frequency direction and bandwidth, and consider vendor metal-artefact options.

Frequently asked questions

Is contrast always needed for prostate MRI?

PI-RADS v2.1 describes DCE as part of multiparametric MRI; some pathways use biparametric MRI (T2 and DWI only) for selected patients. Follow local policy.

Why centre at the symphysis pubis?

The prostate lies just behind and above it, so centring there places the gland at the centre of the coil and magnet.

References and further reading

  • Turkbey B, Rosenkrantz AB, Haider MA, et al. Prostate Imaging Reporting and Data System version 2.1: 2019 update of Prostate Imaging Reporting and Data System version 2. Eur Urol 2019;76(3):340–351.
  • Westbrook C. Handbook of MRI Technique. 4th ed. Wiley-Blackwell; 2014.
  • Westbrook C, Talbot J. MRI in Practice. 5th ed. Wiley-Blackwell; 2018.

Related articles

Last reviewed: 6 November 2026. Educational content only. Positioning, coverage, sequences and parameters are illustrative examples. Always follow your department protocol, the referral question, MRI safety screening and the vendor recommendations for your scanner and coils.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.