MR spectroscopy (MRS) explained: NAA, choline, creatine and lactate

Learning objectives

  • Explain what a proton MR spectrum shows
  • Identify the main brain metabolite peaks and what they mean
  • Choose between short and intermediate echo times
  • Know the practical steps that make or break a spectrum
Single-voxel MR spectroscopy with the voxel position and spectrum
Proton MR spectroscopy: the voxel position shown on three MRI planes (left) and the resulting spectrum (right). Image: Boucard CC et al., PLoS ONE, CC BY 2.5, via Wikimedia Commons.

From image to spectrum

An MR image maps where the water signal is. MR spectroscopy (MRS) instead measures the small signals from other molecules in a chosen volume of tissue (a voxel). Each molecule resonates at a slightly different frequency because of its chemical environment. The frequency is shown on the horizontal axis in parts per million (ppm), so the same spectrum can be compared at any field strength. By convention the axis runs from right to left, with higher ppm on the left.

The main brain metabolites

PeakPosition (ppm)What it reflects
N-acetylaspartate (NAA)2.02Healthy neurons and axons; falls when neurons are lost or damaged
Creatine (Cr)3.03Energy metabolism; fairly stable, often used as a reference
Choline (Cho)3.2Cell membrane turnover; rises with cell proliferation and demyelination
Myo-inositol (mI)3.56A glial marker; raised in gliosis and some dementias
Glutamate + glutamine (Glx)2.1–2.5 and 3.6–3.8Neurotransmitter pool; overlapping peaks
Lactate (Lac)1.33 (doublet)Anaerobic metabolism; normally not visible in brain
Lipids0.9–1.3Necrosis, or contamination from scalp fat

In healthy adult brain, NAA is the tallest peak and Cho is slightly taller than Cr. Hunter’s angle is a quick visual check: a line drawn across the tops of mI, Cho, Cr and NAA rises from left to right.

Proton MR spectrum from occipital grey matter
Proton MR spectrum from occipital grey matter before and after 28 months of treatment in creatine transporter deficiency; the creatine peak is labelled. Image: Nielsen MM et al. 2023, CC BY 4.0, via Wikimedia Commons.

The spectrum above, from a child with creatine transporter deficiency, shows how a missing peak can make the diagnosis on its own.

Short or intermediate echo time?

Echo timeShowsUse
Short (about 30–35 ms)More metabolites, including mI, Glx and lipidsBroad metabolic overview
Intermediate (about 135–144 ms)Fewer, cleaner peaks; lactate doublet points downwardsConfirming lactate and separating it from lipid
Long (about 270–288 ms)Simplest spectrum; lactate points upwardsLess commonly used

The lactate doublet inverts at about 135–144 ms because of J-coupling between its protons. Lipid does not invert, which is how the two are told apart.

Getting a good spectrum

  1. Place the voxel carefully away from bone, scalp fat, air and CSF, which all add unwanted signal or distort the field.
  2. Shim well. A uniform field gives narrow peaks; a poor shim smears them together. Many scanners report the water line width.
  3. Suppress water (for example CHESS), because water is thousands of times stronger than the metabolites.
  4. Use enough averages for adequate SNR, especially for small voxels.
  5. Check quality before reporting: a flat baseline, narrow peaks and no large lipid or water residue.

Common patterns

  • High-grade tumour: Cho up, NAA down, with lipid and lactate in necrotic areas.
  • Pyogenic abscess: amino acids, acetate and succinate in the cavity.
  • Ischaemia and mitochondrial disease: lactate.
  • Canavan disease: very high NAA.
  • Creatine deficiency syndromes: absent or very low Cr.

Key points

  • MRS measures metabolites in a voxel; ppm keeps positions the same at any field.
  • NAA, Cr, Cho, mI and lactate are the core peaks to know.
  • Voxel placement, shimming and water suppression decide the quality.

References

  • Öz G, Alger JR, Barker PB, et al. Clinical proton MR spectroscopy in central nervous system disorders. Radiology 2014;270(3):658–679.

Related articles

Last reviewed: 12 November 2026. Educational content for researchers, clinicians and students. Values are typical examples; local protocols, scanner software and manufacturer guidance take priority.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.