Arterial spin labelling in plain English
Arterial spin labelling (ASL) measures cerebral blood flow without an injected contrast agent. Instead it uses the patient’s own blood as a tracer. This article explains the idea, the main variants and what the numbers do and do not tell you.
Learning objectives
- Explain how magnetically labelled blood works as an endogenous tracer.
- Describe the label and control images and why their difference reflects perfusion.
- Define labelling duration and post-labelling delay and why they matter.
- List the main ASL labelling approaches and typical limitations.

The core idea
Blood flowing towards the brain is magnetically labelled, typically in the neck. After a short wait, the labelled blood has reached the brain tissue and slightly changes the signal. A second image is acquired with no labelling (the control). Subtracting label from control cancels the static tissue and leaves a small signal proportional to blood delivered.
The difference signal is only a small percentage of the total, so ASL typically averages many label-control pairs to improve signal-to-noise ratio.
Timing: labelling duration and post-labelling delay
Two timing parameters shape the signal. Labelling duration is how long blood is labelled. Post-labelling delay (PLD) is the wait between the end of labelling and image acquisition, which lets labelled blood arrive in tissue. Too short a PLD leaves labelled blood in the arteries; too long loses signal as the label decays with T1.

Main labelling approaches
| Approach | How labelling is done | Strengths | Limitations |
|---|---|---|---|
| Pulsed ASL (PASL) | Short pulse labels a slab of blood | Efficient labelling | Lower signal-to-noise ratio in many implementations |
| Continuous ASL (CASL) | Long continuous labelling | High labelling effect | Needs special hardware; high energy deposition |
| Pseudo-continuous ASL (PCASL) | Train of pulses mimics continuous labelling | Good balance of signal and practicality on standard scanners | Sensitive to off-resonance in the labelling plane |
A consensus paper from the ISMRM Perfusion Study Group and the European ASL in Dementia consortium recommends PCASL with a 3D readout for clinical use (Alsop et al., 2015).
What ASL numbers mean
With appropriate modelling and calibration, ASL can be converted to cerebral blood flow in mL/100 g/min. The numbers depend on assumptions such as arterial transit time, labelling efficiency and blood T1. Treat quantitative values as protocol-dependent estimates and compare within the same method.
Limitations and artefacts
- Low signal-to-noise ratio compared with contrast-based methods.
- Prolonged arterial transit time (for example in slow flow) can mimic low perfusion.
- Motion between label and control images corrupts the subtraction.
- Off-resonance and susceptibility effects reduce labelling efficiency.
Key takeaways
- ASL uses magnetically labelled blood as a tracer, no contrast injection.
- Label minus control gives a perfusion-weighted signal.
- Labelling duration and post-labelling delay are critical choices.
- PCASL with 3D readout is the consensus recommendation for clinical use.
Frequently asked questions
Does ASL need a contrast agent?
No. The labelled blood acts as the tracer.
Why is the image noisy?
The perfusion signal is a small fraction of the image signal, so many repetitions are averaged.
Can I compare absolute values between sites?
Only with caution. Different labelling, timing and modelling choices affect the result.
References and further reading
- Alsop DC, et al. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications: a consensus of the ISMRM Perfusion Study Group and the European Consortium for ASL in Dementia. Magn Reson Med. 2015;73:102-116.
Related articles
- ASL timing: post-labelling delay and arterial transit time
- Sodium (²³Na) MRI explained
- Welcome to MRISpin
- From DICOM to BIDS with HeuDiConv
Last reviewed: 4 November 2026. Educational content for researchers, clinicians and students. It does not replace local protocols, manufacturer guidance or clinical judgement.