Diffusion MRI Basics: b-values, DWI and the ADC Map
Learning objectives
- Explain how diffusion gradients make MRI sensitive to water motion
- Describe what the b-value controls
- Calculate an apparent diffusion coefficient (ADC) from two images
- Explain why DWI and ADC are read together

The idea in one paragraph
Water molecules move randomly. A diffusion-weighted sequence adds a pair of strong, balanced gradient pulses around the refocusing pulse (the Stejskal–Tanner design). Spins that stay still are fully rephased; spins that move between the two pulses are not, so they lose phase coherence and signal. The more freely water moves, the more signal is lost.
The b-value
The b-value (s/mm²) summarises how strongly the sequence weights diffusion. It grows with gradient strength, gradient duration and the time between the pulses. b = 0 gives a T2-weighted image with no diffusion weighting; clinical brain DWI commonly uses a high b-value around 1000 s/mm² (illustrative; protocols vary).

The signal equation
In the simplest (mono-exponential) model, signal at b relates to signal at b = 0 by:
S(b) = S0 · exp(−b · ADC)
Rearranging with two acquisitions gives the ADC for each voxel:
ADC = ln(S0 / Sb) / b
Worked example
Suppose a voxel has S0 = 1000 and S1000 = 450 at b = 1000 s/mm². Then ADC = ln(1000/450) / 1000 = ln(2.222)/1000 ≈ 0.799/1000 ≈ 0.80 × 10⁻³ mm²/s.
DWI versus ADC
| Image | What bright means | Watch out for |
|---|---|---|
| DWI (high b) | Less signal lost: more restricted diffusion, or long T2 | T2 shine-through: long-T2 tissue looks bright even without restriction |
| ADC map | Freer diffusion (higher ADC) | Restricted diffusion appears dark |
Because the DWI image still carries T2 weighting, a bright DWI area is confirmed as true restriction only when the ADC map is dark in the same place. See the pathology signal article for clinical context.
Why it matters
- DWI is highly sensitive to early ischaemic change
- Diffusion tensor imaging (DTI) extends the idea to direction-dependent diffusion in white matter
- Diffusion models beyond mono-exponential (for example IVIM) separate perfusion-like effects at low b-values
Key takeaways
- Diffusion gradients convert motion into signal loss
- The b-value sets how strongly diffusion is weighted
- ADC = ln(S0/Sb)/b; read DWI and ADC together
FAQ
Why is it called “apparent” diffusion coefficient?
Tissue is not free water: membranes, cells and perfusion all affect the measurement, so it is an apparent value rather than the true molecular diffusion coefficient.
Does a higher b-value always help?
Higher b increases diffusion contrast but lowers signal-to-noise ratio and can increase distortion, so b is a trade-off.
References
- Stejskal EO, Tanner JE. Spin diffusion measurements: spin echoes in the presence of a time-dependent field gradient. J Chem Phys 1965;42:288-292.
- Le Bihan D, Breton E, Lallemand D, et al. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology 1986;161:401-407.
- Brown RW, Cheng YCN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design, 2nd ed. Wiley; 2014.
Educational disclaimer: values and parameters here are illustrative and vary by scanner, field strength, vendor and protocol. This article is for learning and is not clinical advice.