BodyBody Imaging PlanningMRI Planning

MRCP (Magnetic Resonance Cholangiopancreatography): Protocol, Positioning and Planning

Protocol and Planning

Magnetic resonance cholangiopancreatography (MRCP) uses heavily T2-weighted images to show fluid in the bile ducts and pancreatic duct as bright signal, without contrast and without ionising radiation. This guide follows a full examination: anatomy, indications, safety and preparation, positioning, and then each sequence with its planning and example parameters.

Anatomy of the biliary system

Bile made by the liver drains through the intrahepatic ducts into the right and left hepatic ducts, which join at the hilum to form the common hepatic duct. The cystic duct from the gallbladder joins it to form the common bile duct (CBD). The CBD runs behind the duodenum and through the head of the pancreas, and usually joins the main pancreatic duct (duct of Wirsung) at the ampulla of Vater, which opens into the second part of the duodenum at the major papilla. An accessory pancreatic duct (duct of Santorini) may drain separately through the minor papilla, as in the image below.

Normal MRCP with labelled bile ducts and pancreatic ducts
Normal MRCP, including the accessory pancreatic duct (duct of Santorini). Image: Hellerhoff, CC BY-SA 4.0, via Wikimedia Commons.

Variants matter for reporting and for surgery and endoscopy: an aberrant right posterior hepatic duct, a low or medial cystic duct insertion, and pancreas divisum, where the dorsal pancreatic duct drains through the minor papilla.

Indications

  • Suspected stones in the bile ducts (choledocholithiasis)
  • Biliary obstruction or dilatation of unknown cause, including benign and malignant strictures
  • Primary sclerosing cholangitis and other cholangiopathies
  • Biliary anatomy before surgery, and complications after cholecystectomy or liver transplantation
  • Congenital anomalies such as choledochal cysts and pancreas divisum
  • Chronic pancreatitis and pancreatic duct abnormalities
  • Cystic pancreatic lesions, including follow-up of intraductal papillary mucinous neoplasms (IPMN)

Safety screening

Every patient completes standard MRI safety screening before entering the scanner room. Implants and devices are checked against their MR Conditional labelling for your scanner. Abdominal surgery is common in this group, so ask specifically about surgical clips, stents and biliary or pancreatic drains. Pregnancy is assessed under local policy. See MRI implant and device screening for the full process.

Patient preparation

  • Ask the patient to fast for about 4 hours before the scan. This reduces fluid in the stomach and duodenum, reduces bowel movement and helps the gallbladder to fill.
  • Some departments give a negative oral contrast agent (for example pineapple or blueberry juice, which contain manganese) shortly before scanning to darken fluid in the stomach and duodenum that could overlap the ducts.
  • Explain the breathing commands and practise them before the patient lies down. Consistent breath-holds, usually in expiration, are the most important factor for image quality.
  • Remove all metal and change into a gown. Record the patient’s weight for SAR calculation.
  • Give earplugs and headphones and the call bell.
  • If secretin-enhanced MRCP is requested to assess pancreatic duct function, follow the dedicated local protocol for its preparation and administration.

Positioning

  1. Place the posterior (spine) coil elements on the table and connect any additional body array coil.
  2. Lay the patient supine, feet first or head first according to local practice, with a pillow under the head and a support under the knees.
  3. Place the arms above the head if the patient can keep them there comfortably; otherwise rest them by the sides on pads, away from the abdomen, to avoid wrap and RF contact with the body.
  4. If respiratory triggering is used, place the respiratory bellows (or navigator) on the abdomen where breathing movement is greatest, usually just below the ribs.
  5. Place the anterior body array over the upper abdomen so that it covers the liver and pancreas, and secure it with straps to reduce breathing movement artefact.
  6. Centre the laser a few centimetres below the xiphoid process, at about the level of the lower costal margin, and move the patient to isocentre.
  7. Give the call bell, check comfort, and remind the patient of the breathing commands.

General photos of MRI equipment and workflow; your scanner, coils and room will look different. Images via Wikimedia Commons (credits in each image’s media details).

MRCP protocol and planning

The sequences below make up a typical MRCP examination. Names differ between vendors: HASTE (Siemens) is called SSFSE (GE) or SSh-TSE (Philips); SPACE is called CUBE (GE) or VISTA (Philips). All parameters are illustrative examples for a 1.5 T system and must be adapted to your scanner.

1. Three-plane localiser

Acquire a fast localiser in three planes with the patient breathing gently or in a breath-hold. Check that the liver, gallbladder, bile ducts and pancreas are inside the coil’s sensitive area, and reposition the coil if the signal is low at the top or bottom of the liver.

2. Coronal T2 HASTE, breath-hold

Coronal T2 HASTE planning on an axial abdominal image
Coronal slices planned on an axial image, from the anterior abdominal wall to behind the kidneys. Planning lines added by MRISpin for illustration. Axial and coronal T2 images: Hellerhoff, CC BY-SA 3.0, via Wikimedia Commons.

Planning: plan the coronal slices on the axial localiser, parallel to the left–right axis of the body. Cover from the anterior abdominal wall to behind the kidneys. Check on the sagittal localiser that the stack covers the liver from its dome to below the duodenum. Add a saturation band if needed, and set the phase direction left–right to keep the arms out of the wrap direction where possible.

ParameterExample value (illustrative)
Slice thickness / gap5 mm / 10%
TR / TEabout 1000–1500 ms / 80–90 ms
Field of view350–400 mm
Matrix256 × 256
BreathingOne or two breath-holds in expiration

3. Axial T2 HASTE, breath-hold

Axial T2 HASTE planning on a coronal MRCP image
Axial slices planned on a coronal image, from above the hepatic confluence to below the duodenal papilla. Planning lines added by MRISpin for illustration. MRCP image: Hellerhoff, CC BY-SA 4.0; axial T2: Hellerhoff, CC BY-SA 3.0, via Wikimedia Commons.

Planning: plan the axial slices on the coronal image. Cover from above the hepatic duct confluence (ideally the liver dome) to below the duodenal papilla and the third part of the duodenum. Keep the slices straight on the sagittal localiser.

ParameterExample value (illustrative)
Slice thickness / gap5 mm / 10%
TR / TEabout 1000–1500 ms / 80–90 ms
Field of view350–380 mm
Phase directionAnterior–posterior
BreathingTwo or three breath-holds in expiration

4. Axial T2 HASTE with fat suppression

Planning: copy the axial HASTE geometry so that the two series match slice for slice. Fat suppression makes fluid stand out against fat, which helps with small duct stones, collections and pancreatic changes.

ParameterExample value (illustrative)
Slice thickness / gap5 mm / 10%
Fat suppressionSpectral (SPAIR) or similar
Other settingsAs the axial HASTE

5. Thick-slab single-shot MRCP, coronal oblique

Thick-slab MRCP planning along the common bile duct
A single 40 mm coronal oblique slab placed on the axial image through the pancreatic head and common bile duct, and the resulting MRCP. Planning lines added by MRISpin for illustration. Axial T2: Hellerhoff, CC BY-SA 3.0; MRCP image: Hellerhoff, CC BY-SA 4.0, via Wikimedia Commons.

Planning: on an axial image through the pancreatic head, place a single 40 mm slab through the common bile duct, angled so that it runs along the duct towards the hilum (coronal oblique). Check on the coronal image that the slab includes the hepatic duct confluence and the ampulla. Each slab takes only a few seconds in one breath-hold, so it can be repeated quickly if the patient breathes.

ParameterExample value (illustrative)
Slab thickness40 mm (single slice)
TEvery long, about 700–1000 ms
TRlong (single-shot acquisition)
Field of view300 mm
Matrix384 × 384 or similar
BreathingOne short breath-hold per slab

6. Radial thick-slab MRCP

Radial thick-slab MRCP planning around the common bile duct
Radial thick slabs rotated around the common bile duct on an axial image; each line marks the centre of one slab. Result shown: MRCP with gallbladder and duct stones. Planning lines added by MRISpin for illustration. Images: Hellerhoff, CC BY-SA 3.0, via Wikimedia Commons.

Planning: centre a series of thick slabs on the common bile duct on the axial image and rotate them around it, for example at 15–20° steps across a range of about 120–180°. This gives several views of the ducts from different angles, which helps when duct segments overlap. Many vendors offer a radial planning tool that sets the angles automatically.

ParameterExample value (illustrative)
Slab thickness40 mm per slab
Number of slabsTypically 6–12
TE / TRAs the single thick-slab MRCP
BreathingOne breath-hold per slab or a few slabs per breath-hold, depending on vendor

7. 3D T2 MRCP (SPACE / CUBE / VISTA), respiratory-triggered

3D MRCP volume planning on coronal and axial images
3D respiratory-triggered MRCP volume planned on coronal and checked on axial images, with a 3D MIP result. Planning lines added by MRISpin for illustration. MRCP image: Hellerhoff, CC BY-SA 4.0; axial T2: Hellerhoff, CC BY-SA 3.0; 3D MRCP: Tomáš Vendiš, CC BY 3.0, via Wikimedia Commons.

Planning: plan a coronal (or coronal oblique) 3D volume on the axial and coronal images. Cover the whole biliary tree, from the intrahepatic ducts to the ampulla, and the pancreatic duct to the tail. Check on the axial image that the slab is thick enough to include the CBD and the pancreatic duct, but no thicker than needed, because a thicker slab adds scan time. Place the respiratory trigger or navigator correctly and check that the patient’s breathing trace is regular before starting.

ParameterExample value (illustrative)
Voxel sizeabout 1–1.5 mm (near isotropic)
TEabout 600–700 ms
Fat suppressionYes
BreathingRespiratory-triggered (bellows) or navigator-gated
Scan timeAbout 3–6 minutes, depending on breathing

After the scan, create maximum intensity projections (MIPs) in several rotations and send them with the source images. Small stones and strictures can be hidden on MIPs, so the source images are always reviewed.

Optional sequences

  • Axial T1 in-phase and opposed-phase (Dixon) for liver and pancreas tissue
  • Axial DWI for pancreatic and hepatic lesions
  • Dynamic T1 with gadolinium for masses, strictures and cholangiocarcinoma
  • Secretin-enhanced MRCP for pancreatic duct function

Examples of MRCP findings

MRCP of pancreas divisum
Pancreas divisum on MRCP: (a) gallbladder, (b) common bile duct crossing (c) the long duct of Santorini, (d) short ventral pancreatic duct, (e) right renal collecting system. Image: Hellerhoff, CC BY-SA 3.0, via Wikimedia Commons.

Tips and common pitfalls

  • Breathing is the main cause of poor MRCP. Practise breath-holds, give clear commands and use respiratory triggering for 3D sequences.
  • Bowel fluid in the stomach and duodenum can overlap the ducts on thick slabs; fasting, negative oral contrast and a slightly different slab angle help.
  • Flow voids in the CBD can mimic stones; check on the axial HASTE and the source images.
  • Pneumobilia (air in the ducts after sphincterotomy) and surgical clips also cause signal voids.
  • Overlapping ducts on a single slab can hide a stricture; radial slabs and 3D MIPs at several angles solve this.
  • Ascites is very bright on MRCP and can obscure ducts; adjust the slab and rely on the 3D source images.

Frequently asked questions

Does MRCP need contrast?

No. MRCP uses the natural T2 signal of fluid. Gadolinium is only added when the clinical question needs tissue enhancement, for example for a mass or a stricture.

Why does the patient need to fast?

Fasting reduces fluid and movement in the stomach and duodenum, which can overlap the ducts, and allows the gallbladder to distend.

References and further reading

  • Griffin N, Charles-Edwards G, Grant LA. Magnetic resonance cholangiopancreatography: the ABC of MRCP. Insights Imaging 2012;3(1):11–21. PMC3292642
  • The efficacy of pineapple juice as a negative oral contrast agent in magnetic resonance cholangiopancreatography. S Afr J Radiol. PMC7433279
  • Westbrook C. Handbook of MRI Technique. 4th ed. Wiley-Blackwell; 2014.

Related articles

Last reviewed: 6 November 2026. Educational content only. Positioning, coverage, sequences and parameters are illustrative examples. Always follow your department protocol, the referral question, MRI safety screening and the vendor recommendations for your scanner and coils.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.