MRI gradients and spatial encoding: slice selection, frequency and phase
Learning objectives
- Explain how a magnetic field gradient makes frequency depend on position
- Work out slice thickness from RF bandwidth and gradient strength
- Describe frequency encoding and phase encoding
- Link gradient performance to scan speed, noise and nerve stimulation

Making frequency depend on position
Protons precess at the Larmor frequency, which is proportional to the magnetic field: f = γ̄B, where γ̄ = 42.58 MHz/T for hydrogen. If the field were the same everywhere, every proton would give the same frequency and we could not tell where the signal came from. Three extra coils, the gradient coils (x, y and z), add a small field that changes linearly with position:
f(x) = γ̄ · (B0 + Gx · x)
With a gradient switched on, frequency becomes a label for position. The image is built from three uses of this idea: slice selection, frequency encoding and phase encoding. The gradient direction for each is chosen freely, which is why MRI can image in any plane without moving the patient.
1. Slice selection
During the RF pulse a gradient is applied along the slice direction. Only protons whose frequency falls within the pulse’s bandwidth are excited. The slice thickness is:
slice thickness = RF bandwidth / (γ̄ · G)
gamma_bar = 42.58e6 # Hz per tesla
G = 10e-3 # 10 mT/m
bw = 1000 # 1 kHz RF bandwidth
print(bw / (gamma_bar * G) * 1000, "mm") # ≈ 2.35 mm
Thinner slices need a stronger gradient or a narrower RF bandwidth (a longer pulse). Changing the centre frequency of the pulse moves the slice.
2. Frequency encoding
While the echo is being recorded, a readout gradient is switched on. Each column of the image now emits a slightly different frequency, and the Fourier transform separates them. The receiver bandwidth and the readout gradient together set the field of view in that direction. The fat–water chemical shift appears along this direction as a spatial shift, which is why chemical shift artefact is seen in the frequency-encoding direction.
3. Phase encoding

The third direction is encoded with a brief gradient before the readout. While it is on, spins precess at different rates; when it is switched off, they return to the same frequency but keep a phase difference that depends on position. Each repetition uses a different phase-encoding step, filling one line of k-space. For a conventional 2D sequence:
scan time = TR × number of phase-encoding steps × number of averages
Because phase encoding needs one step per line, it is the slow direction. That is why parallel imaging and partial Fourier reduce phase-encoding steps to save time, and why motion artefacts appear along the phase-encoding direction.
Gradient performance
| Property | Meaning | Why it matters |
|---|---|---|
| Amplitude (mT/m) | Strength of the gradient | Thinner slices, higher resolution, stronger diffusion weighting |
| Slew rate (T/m/s) | How fast the gradient can be switched | Shorter TE and TR, faster EPI, less distortion |
| Duty cycle | How long high gradients can be sustained | Long diffusion and EPI scans heat the gradient coils |
Fast-switching gradients cause the loud knocking noise of MRI (the coils move slightly inside the main field) and can cause peripheral nerve stimulation. Scanners limit gradient switching to stay within safety limits.
References
- Lauterbur PC. Image formation by induced local interactions: examples employing nuclear magnetic resonance. Nature 1973;242:190–191.
- McRobbie DW, Moore EA, Graves MJ, Prince MR. MRI from Picture to Proton. 3rd ed. Cambridge University Press; 2017.