BodyMRI Protocols

Cardiac MRI protocol: cine, mapping, perfusion and late gadolinium enhancement

Learning objectives

  • Describe the building blocks of a standard cardiac MRI examination
  • Plan the main cardiac views from localisers
  • Set up late gadolinium enhancement with an inversion time scout
  • Recognise ischaemic and non-ischaemic patterns of enhancement
Four-chamber late gadolinium enhancement cardiac MRI of a chronic infarct
Four-chamber late gadolinium enhancement image showing a chronic inferior myocardial infarction (bright scar). Image: Doregan, CC BY-SA 4.0, via Wikimedia Commons.

How cardiac MRI is different

The heart moves with every beat and every breath, so cardiac MRI uses ECG gating to acquire data at the same point in the cardiac cycle, and breath-holds (or respiratory navigators) to freeze breathing. Views are planned along the heart’s own axes, not the body’s.

Building blocks of a standard examination

BlockTypical sequenceWhat it gives
LocalisersMulti-plane bSSFP scoutsStarting point for planning the cardiac axes
Cine imagingbSSFP cine: 2-, 3- and 4-chamber views and a short-axis stack (6–8 mm slices, 2–4 mm gaps)Volumes, ejection fraction, wall motion, valves
Oedema / tissue characterisationT2-weighted (e.g. STIR) and/or T2 mapping; native T1 mappingAcute injury, inflammation, diffuse fibrosis, infiltration
Perfusion (if indicated)First-pass gadolinium at stress and restMyocardial ischaemia
Late gadolinium enhancement (LGE)Inversion recovery gradient echo (often PSIR), about 10 minutes after contrastScar and fibrosis
Flow (if indicated)Phase-contrastValve regurgitation, shunts

Based on the SCMR standardised protocols (2020 update). The exact modules depend on the clinical question, such as cardiomyopathy, ischaemia, myocarditis or congenital heart disease.

Cardiac MRI cine image
A cardiac MRI cine frame. Image: Bionerd, CC BY-SA 3.0, via Wikimedia Commons.

Planning the main views

  1. From the axial localiser, plan a vertical long-axis (2-chamber) through the mitral valve and apex.
  2. From the 2-chamber, plan a horizontal long-axis through the apex and the middle of the mitral valve.
  3. From the horizontal long-axis, plan short-axis slices parallel to the mitral valve plane, from base to apex.
  4. From the short-axis, plan the true 4-chamber, 2-chamber and 3-chamber (LV outflow) views.

Late gadolinium enhancement

Gadolinium washes out of normal myocardium but lingers in scar and fibrosis, where the extracellular space is larger. About 10 minutes after injection, an inversion recovery sequence nulls normal myocardium so that scar appears bright. A TI scout (Look-Locker) is run first to find the inversion time where normal myocardium is darkest. Because the gadolinium washes out, the best TI gets longer during the scan. PSIR reconstruction makes the result less sensitive to the exact TI. See inversion recovery and the null point.

Patterns of enhancement

PatternUsually suggests
Subendocardial or transmural, in a coronary territoryMyocardial infarction (ischaemic)
Mid-wall (e.g. septal stripe)Non-ischaemic cardiomyopathy
Subepicardial, often inferolateralMyocarditis
Patchy in the hypertrophied segmentsHypertrophic cardiomyopathy
Basal inferolateral mid-wallTypical of Fabry disease
Cardiac MRI showing left ventricular hypertrophy and fibrosis
Cardiac MRI for left ventricular hypertrophy and fibrosis in Fabry disease and hypertrophic cardiomyopathy. Image: Politei JM, Germain DP et al., CC BY 2.0, via Wikimedia Commons.

Practical tips

  • Check the ECG trace carefully: a clean R wave is essential. Place electrodes after skin preparation, and use vector ECG where available.
  • Practise breath-holds with the patient; end-expiration breath-holds are more reproducible.
  • For arrhythmias, use arrhythmia rejection, real-time cine or free-breathing options.
  • Check implant and device status; many cardiac devices are MR Conditional only under specific settings.

References

  • Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson 2020;22:17.

Related articles

Last reviewed: 7 November 2026. Protocols and parameters are illustrative examples. Local protocols, scanner software, radiologist preference and manufacturer guidance take priority. MRI safety screening applies to every patient.