MRI Physics

BOLD fMRI explained: from blood oxygenation to activation maps

Learning objectives

  • Explain where the BOLD signal comes from
  • Describe the haemodynamic response and its timing
  • Choose sensible echo time and repetition time for BOLD fMRI
  • Recognise the main artefacts and confounds in fMRI
fMRI BOLD activation during finger tapping
BOLD fMRI activation during an index-finger tapping task, analysed with FSL (BET, FAST and FEAT). Image: Zereshk, public domain, via Wikimedia Commons.

Blood as a natural contrast agent

Haemoglobin changes its magnetic properties with oxygen. Oxyhaemoglobin is diamagnetic, close to the surrounding tissue, while deoxyhaemoglobin is paramagnetic. Deoxyhaemoglobin creates tiny field disturbances around vessels, which speed up T2* decay and lower the gradient-echo signal. Ogawa and colleagues described this blood oxygenation level dependent (BOLD) contrast in 1990.

Why active brain becomes brighter

When neurons become active, local blood flow increases more than oxygen use. The extra oxygenated blood washes out deoxyhaemoglobin, T2* lengthens slightly, and the signal rises. The change is small: typically a few per cent at 3 T, and larger at 7 T.

The haemodynamic response

The BOLD response is slow and delayed. After a brief event it peaks after about 5–6 seconds and is often followed by a small undershoot before returning to baseline. Analysis software models this with a haemodynamic response function (HRF). A common choice is the double-gamma function:

import numpy as np
from scipy.stats import gamma

t = np.arange(0, 32, 0.1)                  # seconds
hrf = gamma.pdf(t, 6) - gamma.pdf(t, 16) / 6
hrf /= hrf.max()
print(t[hrf.argmax()])   # peak at 5.0 s
print(t[hrf.argmin()])   # undershoot minimum at about 15.7 s

In a GLM analysis, each condition’s timing is convolved with this function to predict the signal, and the fit tells you where the brain followed the task. See FSL design files explained for how this is set up in FSL.

fMRI BOLD activation in an emotional Stroop task
BOLD activation for incongruent compared with congruent blocks in an emotional Stroop task. Image: Ovaysikia S, Tahir KA, Chan JL, DeSouza JFX, CC BY 2.5, via Wikimedia Commons.

Acquisition: gradient-echo EPI

BOLD fMRI almost always uses single-shot gradient-echo echo-planar imaging (EPI), which acquires a whole slice in tens of milliseconds. The echo time is set close to the T2* of grey matter to maximise BOLD sensitivity.

ParameterTypical valueNote
TE≈ 30 ms at 3 T (longer at 1.5 T, shorter at 7 T)Close to grey matter T2*
TR≈ 0.7–2 sShorter with simultaneous multi-slice (multiband)
Voxel size≈ 2–3 mm isotropicSmaller at 7 T
VolumesHundreds per runDepends on the task design

Artefacts and confounds

  • Distortion and signal loss near the sinuses and ear canals (orbitofrontal cortex, inferior temporal lobes). Field maps or reversed phase-encoding scans are used for correction.
  • Head motion, which can create false activation, especially if it is linked to the task.
  • Physiological noise from breathing and heartbeat.
  • Vascular origin. BOLD measures a blood response, not neurons directly, and large draining veins can displace activation.

References

  • Ogawa S, Lee TM, Kay AR, Tank DW. Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc Natl Acad Sci USA 1990;87(24):9868–9872.
  • Logothetis NK. What we can do and what we cannot do with fMRI. Nature 2008;453:869–878.

Related articles

Last reviewed: 7 November 2026. Educational content for researchers, clinicians and students. Values are typical examples; local protocols, scanner software and manufacturer guidance take priority.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles

Back to top button