MRI Slice Planning Basics: Orientation, Coverage and Angulation
Learning objectives
- Describe the main geometry parameters in a slice prescription
- Explain trade-offs between coverage, resolution and time
- Understand why angulation to landmarks improves comparability

Geometry parameters
| Parameter | Meaning | Trade-off |
|---|---|---|
| Field of view (FOV) | Size of the imaged area | Larger FOV allows coverage but lowers resolution for a given matrix |
| Matrix | Number of pixels across the FOV | More pixels improve detail but cost time and SNR |
| Slice thickness | Through-plane resolution | Thinner slices have less SNR and partial volume |
| Slice gap | Space between slices | Gaps save time but can miss small lesions |
| Angulation | Tilt relative to the scanner axes | Aligns slices to anatomy for reproducibility |
Practical principles
- Plan from a localiser (scout) image in at least two planes
- Cover the whole target region plus a margin
- Align to a stable landmark so follow-up scans are comparable
- Set phase-encoding direction to keep artefacts away from the region of interest
- Add saturation bands when flow or motion from adjacent regions causes artefact
In-plane pixel size is simply FOV divided by matrix. A 240 mm FOV with a 256 matrix gives about 0.94 mm pixels (illustrative numbers).
Key takeaways
- Planning is a trade-off between coverage, resolution, SNR and time
- Consistent angulation makes comparisons over time meaningful
- Always follow the local site protocol; the numbers here are illustrative
FAQ
What is a scout or localiser?
A fast low-resolution scan used to plan the real acquisitions.
Why does phase-encoding direction matter?
Motion and aliasing artefacts propagate along that direction, so it is chosen to protect the anatomy of interest.
References and further reading
- Brown RW, Cheng YCN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design, 2nd ed. Wiley; 2014.
Educational disclaimer: this article is for learning. Parameters and appearances described are illustrative and vary by scanner, field strength, vendor and protocol. It is not medical advice and does not replace local protocols or clinical judgement.