Body Imaging PlanningMRI Planning

MRI Breast: Protocol, Positioning and Planning

This guide covers a breast MRI from start to finish: anatomy and landmarks, indications, safety and preparation, prone positioning from coil to first image, and each sequence with its planning image and example parameters.

Anatomy and planning landmarks

Each breast lies on the pectoralis major muscle and extends from about the second to the sixth rib, with an axillary tail towards the armpit. It is made of glandular (fibroglandular) tissue and fat, and the amount of each varies widely between women. The axillary lymph nodes are included in breast MRI because they are part of staging.

For planning, the field of view must include both breasts completely, both axillae, and the chest wall behind the breasts. Coverage runs from above the axillary tail to below the inframammary folds.

Indications

  • High-risk screening (for example BRCA carriers or strong family history), according to national guidance
  • Local staging of newly diagnosed breast cancer, when recommended
  • Assessing response to neoadjuvant chemotherapy
  • Problem-solving after inconclusive mammography or ultrasound
  • Implant integrity (dedicated silicone protocol)

Safety screening

Every patient completes standard MRI safety screening before entering the scanner room, and any implant or device is checked against its MR Conditional labelling for your scanner. Ask about tissue expanders with magnetic ports, which are often MR Unsafe, and about breast implants and their type. See MRI implant and device screening for the full process.

Patient preparation

  • Check the indication and previous imaging, and screen renal function and allergies for gadolinium
  • For premenopausal patients, many guidelines advise scheduling between about day 7 and day 14 of the menstrual cycle to reduce background enhancement (local protocol)
  • Place an IV cannula before positioning, usually in the arm that will be most comfortable, with extension tubing long enough for the prone position
  • Explain the importance of staying completely still between the pre- and post-contrast series
  • Remove all metal and change into a gown; record the patient’s weight

Positioning

StepTypical approach
PositionProne, head first, on the dedicated breast coil
BreastsHanging centrally in the coil openings, without skin folds; tissue gently pulled forward from the chest wall
ArmsAbove the head or by the sides, depending on coil design and comfort
CentringCentre of the breast coil openings (nipple level)
ComfortPads for the head, chest and feet; a comfortable patient moves less during the dynamic series

Step by step: from coil to first image

  1. Check the request and screening: identity, indication, menstrual cycle timing if relevant, contrast and renal function.
  2. Prepare the breast coil: clean it, add the padding and check that the cable is connected.
  3. Place the IV cannula and connect the extension line before the patient lies down.
  4. Help the patient lie prone on the coil with the breasts hanging centrally in the openings; use the face rest for the head.
  5. Gently pull each breast forward so that tissue near the chest wall is inside the coil and there are no skin folds; use the compression plates lightly if the system has them.
  6. Position the arms above the head or by the sides according to the coil design, and support the feet and shins.
  7. Give the call bell and earplugs or headphones, and explain the breath and movement instructions.
  8. Switch on the laser and landmark at the centre of the coil openings, then move the table to isocentre.
  9. Register the patient on the console (prone, head first) and check that the coil elements are detected.
  10. Run the localiser and check that both breasts and axillae are covered and that tissue is not outside the coil.

General photos of MRI equipment and workflow; your scanner, coils and room will look different. Images via Wikimedia Commons (credits in each image’s media details).

Breast MRI protocol and planning

The sequences below make up a typical examination. Sequence names differ between vendors (for example TSE/FSE, HASTE/SSFSE, SPACE/CUBE/VISTA). All parameters are illustrative examples, mostly for 1.5 T, and must be adapted to your scanner, coil and local protocol.

1. Three-plane localiser

Planning: Check that both breasts, including the chest wall and axillae, are inside the coil, and that there are no skin folds.

2. Axial T1 without fat suppression

Breast MRI bilateral axial field of view planning
Bilateral axial field of view covering both breasts and both axillae, shown on an axial T1 image of a prone breast MRI. Images: Nevit Dilmen, CC BY-SA 3.0 (T1 and T2 channels of a colour composite shown in grey scale), via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Plan on the sagittal and coronal localisers. The axial field of view covers both breasts and both axillae, from above the axillary tail to below the inframammary folds. Set the phase direction to avoid heart and breathing motion crossing the breasts (usually left–right for axial).

ParameterExample value (illustrative)
Slice thickness / gap3–4 mm / 0–10%
TR / TEabout 500–700 ms / 10 ms
Field of viewabout 320–360 mm (both breasts)

3. Axial T2 or STIR

Planning: Copy the axial geometry. Shows cysts, oedema and some fibroadenomas.

ParameterExample value (illustrative)
Slice thickness3–4 mm
TR / TEabout 4000–6000 ms / 70–90 ms (T2)
Fat suppressionSpectral (or STIR)

4. Axial DWI with ADC map

Planning: Copy the axial coverage.

ParameterExample value (illustrative)
TechniqueSingle-shot EPI, fat-suppressed
b-valuesfor example 0–50 and 800 s/mm²
Slice thickness4 mm

5. Axial 3D T1 fat-suppressed dynamic

Planning: Copy the axial coverage. Acquire one pre-contrast and several post-contrast acquisitions after gadolinium without moving the patient. Create subtraction images (post minus pre) and, if needed, maximum intensity projections.

ParameterExample value (illustrative)
Technique3D spoiled gradient echo with fat suppression
Slice thickness1–1.5 mm (near isotropic)
Temporal resolutionabout 60–90 s per phase
Phases1 pre-contrast and about 4–6 post-contrast

6. Sagittal T1 or T2 (unilateral, some sites)

Breast MRI sagittal slice planning over each breast
Sagittal slice stacks planned over each breast on an axial image, from the medial to the lateral edge of the breast. Images: Nevit Dilmen, CC BY-SA 3.0 (T1 and T2 channels of a colour composite shown in grey scale), via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Plan on the axial image over each breast, from its medial to its lateral edge, parallel to the midline.

7. Silicone-specific sequences (implant studies only)

Planning: Use the vendor silicone-only or water-suppressed sequences in two planes, covering both implants.

Tips and common pitfalls

  • Skin folds and tissue left outside the coil cause missed areas; spend time on positioning.
  • Any movement between pre- and post-contrast images ruins subtraction; make the patient comfortable first.
  • Shim over both breasts for uniform fat suppression; poor shimming causes bright fat in one breast.
  • Heart and breathing motion can create artefact across the breasts; choose the phase direction carefully.

Frequently asked questions

Why is breast MRI performed prone?

Hanging the breasts in the coil separates the tissue from the chest wall, reduces breathing movement and improves image quality.

Why do timings in the menstrual cycle matter?

Background enhancement of normal breast tissue changes with the cycle and is lowest in the second week, which reduces false positives.

References and further reading

  • Mann RM, Kuhl CK, Moy L. Contrast-enhanced MRI for breast cancer screening. J Magn Reson Imaging 2019. PMC6767440
  • Westbrook C. Handbook of MRI Technique. 4th ed. Wiley-Blackwell; 2014.
  • Westbrook C, Talbot J. MRI in Practice. 5th ed. Wiley-Blackwell; 2018.

Related articles

Last reviewed: 6 November 2026. Educational content only. Positioning, coverage, sequences and parameters are illustrative examples. Always follow your department protocol, the referral question, MRI safety screening and the vendor recommendations for your scanner and coils.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.