MRI Protocols

MRI protocol for multiple sclerosis: brain and spinal cord sequences explained

Learning objectives

  • List the core brain and spinal cord sequences for a multiple sclerosis (MS) diagnostic scan
  • Explain why 3D FLAIR is the backbone of MS imaging
  • Know when gadolinium is and is not routinely needed
  • Recognise the extra sequences used for cortical lesions and the central vein sign

What the scan has to show

MRI supports the diagnosis of MS by showing lesions that are disseminated in space (in typical regions of the central nervous system) and, in some situations, disseminated in time. The typical regions are periventricular, cortical or juxtacortical, infratentorial and spinal cord. The 2024 revision of the McDonald criteria added the optic nerve as a fifth region and recognised the central vein sign and paramagnetic rim lesions as supportive MRI findings. The full rules, including when dissemination in time is required, should be read in the original criteria.

Because the diagnosis depends on comparing scans, the most important protocol feature is consistency: the same field strength, slice orientation, slice thickness and sequences at every visit where possible.

Brain protocol at diagnosis

SequencePurposeExample settings (illustrative)
3D T2-FLAIR (sagittal acquisition)Main sequence for counting and locating lesions; reformats in all planesAbout 1 mm isotropic
Axial T2 (or PD/T2)Infratentorial lesions, which FLAIR can missAbout 3 mm, no gap
3D T1 before and after gadoliniumEnhancing (active) lesions at diagnosis; black holes; atrophy measuresAbout 1 mm isotropic; post-contrast after a short delay
DWIDifferential diagnosis, e.g. acute infarctAxial, b0 and b1000
Optional: DIR or PSIRCortical lesionsPer vendor; best at 3 T
Optional: SWI or T2* (3D EPI)Central vein sign; paramagnetic rim lesionsHigh-resolution susceptibility imaging

Settings are illustrative examples of the kind recommended in the 2021 MAGNIMS–CMSC–NAIMS consensus; they are not a validated protocol for any specific scanner. Field strength of at least 1.5 T is recommended, with 3 T preferred where available.

Axial T2-weighted brain MRI in multiple sclerosis
Axial T2-weighted brain MRI in multiple sclerosis. Image: James Heilman, MD, CC BY-SA 4.0, via Wikimedia Commons.

Spinal cord protocol at diagnosis

Spinal cord imaging is recommended at diagnosis because cord lesions add regions for dissemination in space and help with the differential diagnosis. The consensus recommends at least two sagittal sequences of the cervical cord (and ideally the whole cord), chosen from T2, PD and STIR, plus axial T2 or gradient-echo images through lesions. A post-contrast sagittal T1 can be added when the brain scan is also contrast-enhanced.

SequencePurpose
Sagittal T2 TSELesion detection and cord shape
Sagittal PD or STIRSecond contrast to confirm lesions; STIR is sensitive in the cervical cord
Axial T2 or T2* through lesionsConfirms lesions and shows their position in the cord cross-section
Sagittal T1 after contrast (optional)Active cord lesions

Follow-up and monitoring

  • For routine monitoring of treated patients, the 2021 consensus advises that gadolinium is not routinely required. New or enlarging T2 lesions on a repeat 3D FLAIR, compared with the previous scan, are the main marker of new disease activity.
  • Contrast remains useful in specific situations, for example to confirm activity at baseline before treatment or when the clinical picture is unclear.
  • Repeat spinal cord MRI is not routinely needed for monitoring, but is considered for unexpected clinical change or for prognosis in some patients.
  • Patients at higher risk of progressive multifocal leukoencephalopathy (PML) on some treatments may have an abbreviated brain protocol (FLAIR, T2 and DWI) at shorter intervals, according to local guidance.

Practical tips for radiographers and technologists

  • Plan the 3D FLAIR and 3D T1 in the same way at every visit and save the planning as a protocol so that follow-up scans match.
  • Check that the FLAIR covers the whole brain including the brainstem and cerebellum.
  • Use the same post-contrast delay at every contrast scan, because enhancement depends on timing.
  • If motion degrades the 3D FLAIR, repeat it before moving on: it is the sequence the report depends on.
  • Record the gadolinium agent and dose in the examination record.

Frequently asked questions

Why is 3D FLAIR preferred over 2D FLAIR?

Thin isotropic voxels detect small lesions, reduce partial-volume effects and can be reformatted in any plane, which makes comparison between visits more reliable.

Are DIR and PSIR required for the diagnosis?

No. They improve detection of cortical lesions, as in the images above, but are optional and depend on the scanner and local expertise.

References and further reading

  • Wattjes MP, Ciccarelli O, Reich DS, et al. 2021 MAGNIMS–CMSC–NAIMS consensus recommendations on the use of MRI in patients with multiple sclerosis. Lancet Neurol 2021;20(8):653–670. PubMed
  • Montalban X, et al. Diagnosis of multiple sclerosis: 2024 revisions of the McDonald criteria. Lancet Neurol 2025;24(10):850–865.
  • Favaretto A, Poggiali D, Lazzarotto A, et al. The parallel analysis of phase sensitive inversion recovery (PSIR) and double inversion recovery (DIR) images significantly improves the detection of cortical lesions in multiple sclerosis (MS) since clinical onset. PLoS ONE 2015. PMC4444021

Related articles

Last reviewed: 6 November 2026. Educational content only. Example parameters are illustrative and vary by scanner, vendor and local protocol. It does not replace your institution’s policies, the manufacturer’s instructions, or the judgement of your MR safety expert and reporting radiologist.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles