MRI PlanningNeuro Planning

MRI Brain: Protocol, Positioning and Planning

This guide covers a routine brain MRI from start to finish: the anatomy and landmarks you plan on, indications, safety and preparation, positioning from coil to first image, and then each sequence with its planning image and example parameters.

Anatomy and planning landmarks

The brain is planned on a few reliable midline landmarks seen on the sagittal localiser. The anterior commissure (AC) and posterior commissure (PC) are small white-matter bundles in front of and behind the third ventricle; the line joining them (AC–PC line) is the standard reference for axial slices. When the commissures are hard to see, the line along the lower edge of the genu and splenium of the corpus callosum (subcallosal line) is a practical substitute.

On the coronal and axial localisers, the midline (falx, third ventricle, septum pellucidum) is used to correct head tilt and rotation. Coverage runs from the foramen magnum to the vertex, and from the frontal to the occipital poles. For more detail see MRI brain anatomy and imaging planes.

Indications

  • Headache with red-flag features, new neurological symptoms or seizures
  • Suspected stroke, transient ischaemic attack or vascular lesion
  • Suspected or known tumour, metastases or infection
  • Demyelinating disease such as multiple sclerosis
  • Dementia and cognitive decline
  • Follow-up after surgery, radiotherapy or treatment

Safety screening

Every patient completes standard MRI safety screening before entering the scanner room, and any implant or device is checked against its MR Conditional labelling for your scanner. Ask specifically about aneurysm clips, shunts with programmable valves, cochlear implants, neurostimulators and metal in the eyes. See MRI implant and device screening for the full process.

Patient preparation

  • Remove hairpins, hair accessories, hearing aids, dentures with metal, glasses and make-up containing metal
  • Explain the scan length and noise; give earplugs and headphones
  • Check whether contrast is requested and that renal function and allergy screening are complete; place an IV cannula if needed
  • Ask the patient to keep their head still and their eyes closed during scans, which reduces eye-movement ghosting
  • Record the patient’s weight for SAR calculation

Positioning

StepTypical approach
PositionSupine, head first
CoilMultichannel head coil or head–neck coil
AlignmentInterpupillary line horizontal; midsagittal plane straight along the table
CentringLaser at the nasion or glabella
ImmobilisationFoam pads either side of the head; knee support for comfort

Step by step: from coil to first image

  1. Check the request and screening: identity, indication, contrast needs and completed MRI safety screening.
  2. Prepare the room: clean the table, select the head coil, and have pads, a knee wedge, earplugs or headphones and the call bell ready.
  3. Connect the posterior part of the head coil and place the head cushion.
  4. Lay the patient supine, head first, with the head in the posterior coil and the shoulders touching the coil base.
  5. Align the head: interpupillary line horizontal and midsagittal plane straight along the table.
  6. Pad both sides of the head firmly but comfortably; place a support under the knees.
  7. Attach the anterior coil and lock it, checking it does not press on the nose or eyes; place the mirror if used.
  8. Give the call bell and fit earplugs or headphones.
  9. Switch on the laser, landmark at the nasion or glabella and move the table to isocentre.
  10. Register the patient on the console (details, weight, orientation, coil) and check that the coil elements are detected.
  11. Run the three-plane localiser and check it: head centred, no tilt or rotation, good coil signal. Reposition and repeat if needed.

General photos of MRI equipment and workflow; your scanner, coils and room will look different. Images via Wikimedia Commons (credits in each image’s media details).

Brain MRI protocol and planning

The sequences below make up a typical examination. Sequence names differ between vendors (for example TSE/FSE, HASTE/SSFSE, SPACE/CUBE/VISTA). All parameters are illustrative examples, mostly for 1.5 T, and must be adapted to your scanner, coil and local protocol.

1. Three-plane localiser

Planning: Acquire a fast localiser in three planes. Use the coronal and axial images to check that the head is straight; if it is tilted or rotated, reposition rather than relying only on angled slices.

2. Sagittal T1 (or 3D T1)

Brain MRI sagittal slice planning on axial and coronal localisers
Sagittal slices planned parallel to the midline (falx) on the axial and coronal localisers, covering both temporal lobes. Localiser images: Osirinisaboon, CC BY-SA 4.0, via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Plan on the axial and coronal localisers, parallel to the midline (falx). Cover the whole brain from one temporal lobe to the other. The mid-sagittal image is then used to angle the axial and coronal slices.

ParameterExample value (illustrative)
Slice thickness / gap5 mm / 10% (or 1 mm isotropic 3D)
TR / TEabout 500–600 ms / 10–15 ms (spin echo)
Field of view230–240 mm
Matrix256 × 224

3. Axial T2 TSE

Brain MRI axial slice planning on sagittal and coronal localisers
Axial slices planned parallel to the AC–PC line on the sagittal localiser and straightened on the coronal localiser, from the foramen magnum to the vertex. Localiser images: Osirinisaboon, CC BY-SA 4.0, via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Plan on the mid-sagittal image parallel to the AC–PC line (or the subcallosal line), and straighten the block on the coronal image so that it is perpendicular to the midline. Cover from the foramen magnum to the vertex. Use the same geometry for all axial sequences so that they match slice for slice.

ParameterExample value (illustrative)
Slice thickness / gap5 mm / 10%
TR / TEabout 4000–6000 ms / 90–110 ms
Echo train length10–16
Field of view220–230 mm
Matrix384 × 300 or similar
Phase directionLeft–right

4. Axial FLAIR

Planning: Copy the axial T2 geometry.

FLAIR suppresses CSF so that lesions next to the ventricles and in the cortex stand out.

ParameterExample value (illustrative)
Slice thickness / gap5 mm / 10%
TR / TEabout 8000–9000 ms / 100–120 ms
Inversion timeabout 2000–2500 ms at 1.5 T
Field of view220–230 mm

5. Axial DWI with ADC map

Planning: Copy the axial geometry. Check that the ADC map is generated and sent with the images.

ParameterExample value (illustrative)
TechniqueSingle-shot spin-echo EPI
b-values0 and 1000 s/mm²
Slice thickness5 mm
TR / TEabout 3000–5000 ms / minimum
Matrix128–192
Directions3 (trace)

6. Axial SWI or T2* gradient echo

Planning: Copy the axial geometry, or use a 3D slab covering the same range.

Shows blood products, microbleeds, calcification and veins.

ParameterExample value (illustrative)
Technique3D gradient echo (SWI) or 2D T2* GRE
TEabout 40 ms at 1.5 T (about 20 ms at 3 T) for SWI
Slice thicknessabout 2 mm (SWI) or 5 mm (2D T2*)
Flip angleabout 15–20°

7. Coronal T2 or FLAIR

Brain MRI coronal slice planning on sagittal and axial localisers
Coronal slices planned perpendicular to the AC–PC line on the sagittal localiser and straight across the midline on the axial localiser, from the frontal to the occipital pole. Localiser images: Osirinisaboon, CC BY-SA 4.0, via Wikimedia Commons. Planning lines added by MRISpin for illustration.

Planning: Plan on the mid-sagittal image perpendicular to the AC–PC line, and straighten on the axial image so that the slices run straight across the midline. Cover from the frontal to the occipital pole. For epilepsy, angle the coronal slices perpendicular to the long axis of the hippocampus and follow the dedicated epilepsy protocol.

ParameterExample value (illustrative)
Slice thickness / gap4–5 mm / 10%
TR / TEas axial T2 or FLAIR
Field of view220–230 mm

8. 3D T1 after gadolinium (if requested)

Planning: Plan a sagittal or axial 3D block covering the whole brain. Thin isotropic voxels allow reformats in any plane.

ParameterExample value (illustrative)
Technique3D gradient echo (e.g. MPRAGE) or 3D spin echo
Voxel sizeabout 1 mm isotropic
ContrastGadolinium at the dose set by local protocol
TimingAfter a short delay; keep the same delay at follow-up

Optional sequences

  • MR angiography (TOF) for vascular questions; see Circle of Willis on MRI
  • Perfusion (DSC or ASL) for tumours and stroke
  • Thin coronal T2 for the pituitary or hippocampi, and thin axial T2 for the internal auditory meati

Tips and common pitfalls

  • A tilted head makes left–right comparison unreliable; correct it at positioning.
  • Use the same AC–PC angulation at follow-up so that scans are comparable.
  • Eye-movement ghosts run along the phase-encoding direction; ask the patient to close their eyes and keep still.
  • A field of view of about 220–240 mm usually avoids wrap.
  • Run DWI early in unwell patients so that the key images are acquired first.

Frequently asked questions

Why angle to the AC–PC line?

It is a reproducible internal landmark, so axial slices match between patients, visits and brain atlases.

When are thin coronal slices needed?

For epilepsy, the pituitary gland and the internal auditory meati, which have dedicated protocols.

References and further reading

  • Westbrook C. Handbook of MRI Technique. 4th ed. Wiley-Blackwell; 2014.
  • Westbrook C, Talbot J. MRI in Practice. 5th ed. Wiley-Blackwell; 2018.
  • Möller TB, Reif E. Pocket Atlas of Sectional Anatomy, Volume I: Head and Neck. Thieme.

Related articles

Last reviewed: 6 November 2026. Educational content only. Positioning, coverage, sequences and parameters are illustrative examples. Always follow your department protocol, the referral question, MRI safety screening and the vendor recommendations for your scanner and coils.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles

Back to top button