Measuring Tissue Sodium Concentration with ²³Na MRI
Learning objectives
- Explain why sodium MRI needs ultra-short echo times
- Describe how reference phantoms calibrate signal to concentration
- List the main sources of error in tissue sodium concentration (TSC)

Why sodium is hard to image
Sodium-23 gives far less signal than hydrogen: its tissue concentration is about a thousand times lower and its MR sensitivity is lower too. In tissue its transverse relaxation is bi-exponential, with a large fast component of under a millisecond. Conventional echo times would lose most of the signal, so sodium imaging uses ultra-short or zero echo time readouts such as radial or twisted-projection trajectories. For an introduction see sodium MRI explained.

From signal to concentration
Sodium images are usually calibrated with vials of known concentration placed in the field of view. After corrections, tissue concentration is estimated by proportion:
TSC_tissue ≈ C_ref × (S_tissue / S_ref) × corrections
Worked example
If a 50 mmol/L reference vial gives signal 1.00 and a brain region gives 0.80 after corrections, TSC ≈ 50 × 0.80 = 40 mmol/L (illustrative).
Corrections that matter
| Source of error | Why | Typical mitigation |
|---|---|---|
| B1 transmit and receive non-uniformity | Signal varies across the coil, especially at 7 T | B1 mapping and correction |
| Relaxation differences | Tissue and phantom T1/T2* differ | Short TE, long TR, or relaxation correction |
| Partial volume | Voxels are large (often several mm) | Partial-volume correction, careful interpretation near CSF |
| Low SNR | Weak signal | Higher field, dedicated sodium coils, averaging |
Why TSC is interesting
Healthy cells keep intracellular sodium low (around 10–15 mmol/L) against a high extracellular level (around 140 mmol/L) using energy-dependent pumps. Changes in cell integrity, energy metabolism or extracellular volume can change measured TSC, which is why sodium MRI is studied in neurological conditions.
Key takeaways
- Sodium needs ultra-short TE to capture the fast-decaying signal
- TSC is calibrated against reference phantoms
- B1, relaxation and partial-volume corrections strongly affect accuracy
FAQ
Does sodium MRI need special hardware?
Yes. It needs a scanner with multinuclear capability and a coil tuned to the sodium frequency (about 11.26 MHz/T).
Why is 7 T attractive for sodium?
Signal-to-noise increases with field strength, which allows finer resolution or shorter scans, though B1 non-uniformity becomes harder to manage.
References
- Madelin G, Regatte RR. Biomedical applications of sodium MRI in vivo. J Magn Reson Imaging 2013;38:511-529.
- Thulborn KR. Quantitative sodium MR imaging: a review of its evolving role in medicine. NeuroImage 2018;168:250-268.
Educational disclaimer: parameter values are illustrative and depend on field strength, sequence, population and site protocol. This article is for learning and is not clinical advice.