Inversion recovery in MRI: FLAIR, STIR, DIR and the null point
Learning objectives
- Describe what an inversion pulse does to longitudinal magnetisation
- Calculate the null point for a tissue from its T1
- Explain how FLAIR, STIR, DIR, PSIR and MPRAGE use inversion recovery
- Recognise the limits of each technique

The inversion pulse
An inversion recovery (IR) sequence starts with a 180° pulse that flips the longitudinal magnetisation from +M0 to −M0. Each tissue then recovers back towards +M0 at a rate set by its T1. After a chosen inversion time (TI), a normal imaging sequence (spin echo, fast spin echo or gradient echo) reads out whatever magnetisation is present.
For a long repetition time, longitudinal magnetisation during recovery follows:
Mz(TI) = M0 · (1 − 2·e^(−TI/T1))
Every tissue passes through zero on its way back. If the readout happens exactly at that moment, the tissue gives no signal. This is the null point:
TI_null = T1 · ln 2 ≈ 0.69 · T1 (long TR)
TI_null = T1 · ln( 2 / (1 + e^(−TR/T1)) ) (finite TR)
The second form matters when TR is not much longer than T1, as with CSF, because the magnetisation has not fully recovered before the next inversion.
Worked example in Python
import math
def ti_null(t1, tr=None):
if tr is None: # very long TR
return t1 * math.log(2)
return t1 * math.log(2 / (1 + math.exp(-tr / t1)))
print(round(ti_null(4000))) # CSF, long TR -> 2773 ms
print(round(ti_null(4000, 9000))) # CSF, TR 9000 ms -> 2372 ms
print(round(ti_null(260, 3000))) # fat at 1.5 T -> 180 ms
T1 values are approximate and vary between studies and field strengths, which is why vendors provide tested default TIs. CSF T1 is around 4 seconds; fat is roughly 250–290 ms at 1.5 T and longer at 3 T.
FLAIR: nulling CSF
Fluid-attenuated inversion recovery uses a long TI (typically around 2,000–2,800 ms, depending on TR and field strength) to null free water, usually combined with a long-TE fast spin-echo readout. CSF becomes dark while oedema, demyelination and gliosis stay bright, so lesions next to the ventricles and in the cortex become far easier to see. FLAIR is a core sequence for multiple sclerosis, epilepsy, infection and many other brain conditions.
Pitfalls include CSF flow artefacts (bright signal in the basal cisterns and around the fourth ventricle), and high CSF signal when the subject has had supplemental oxygen or contrast, or when CSF protein is high.
STIR: nulling fat
Short tau inversion recovery uses a short TI to null fat. Because it does not rely on the fat resonance frequency, it suppresses fat evenly even near metal and across large fields of view. It is very sensitive to bone marrow oedema, fractures and soft-tissue inflammation.

STIR nulls everything with a T1 close to that of fat, including tissue whose T1 has been shortened by gadolinium. It is therefore not used after contrast.

Double inversion recovery (DIR)
DIR uses two inversion pulses with two inversion times so that both CSF and white matter are nulled. Grey matter and lesions remain, which helps to detect cortical lesions in multiple sclerosis. The price is low SNR and long scan time.
Magnitude versus phase-sensitive reconstruction (PSIR)
A standard magnitude image cannot tell negative magnetisation from positive, so a tissue that is still negative appears bright. Phase-sensitive inversion recovery keeps the sign, giving a wider dynamic range and less dependence on the exact TI. PSIR is used for grey–white contrast in the brain and for late gadolinium enhancement in cardiac MRI.
MPRAGE and other T1-weighted IR
3D T1-weighted gradient-echo sequences such as MPRAGE use an inversion pulse before each block of readouts to strengthen T1 contrast. Here the goal is not to null a tissue but to choose a TI where grey and white matter differ most. This is the standard sequence for brain volumetry and FreeSurfer analysis.
Summary
| Technique | Tissue nulled | Typical use |
|---|---|---|
| FLAIR | CSF (long TI) | Lesions near CSF: MS, epilepsy, infection |
| STIR | Fat (short TI) | Marrow oedema, fractures, large field of view, near metal |
| DIR | CSF and white matter | Cortical MS lesions |
| PSIR | Depends on TI; sign preserved | Grey–white contrast, cardiac scar imaging |
| MPRAGE | None (TI chosen for contrast) | 3D T1 for volumetry |
References
- Bydder GM, Young IR. MR imaging: clinical use of the inversion recovery sequence. J Comput Assist Tomogr 1985;9(4):659–675.
- Hajnal JV, Bryant DJ, Kasuboski L, et al. Use of fluid attenuated inversion recovery (FLAIR) pulse sequences in MRI of the brain. J Comput Assist Tomogr 1992;16(6):841–844.
Related articles
- Fat suppression: STIR, fat saturation, SPAIR and Dixon
- T1 and T2 relaxation explained
- MRI protocol for multiple sclerosis
Last reviewed: 7 November 2026. Educational content for researchers, clinicians and students. Values are typical examples; local protocols, scanner software and manufacturer guidance take priority.