Phase-contrast MRI: measuring blood and CSF flow with VENC
Learning objectives
- Explain how moving spins gain phase in a bipolar gradient
- Choose a sensible velocity encoding (VENC)
- Recognise velocity aliasing
- Know the main clinical uses of phase-contrast MRI

Phase as a speedometer
A bipolar gradient is a gradient pulse followed by an equal pulse of opposite sign. Stationary spins gain phase in the first half and lose exactly the same amount in the second, ending with no net phase. Spins that move along the gradient direction sit in a different field during each half, so they finish with a net phase proportional to their velocity.
Phase-contrast (PC) MRI acquires two scans with different bipolar gradients and subtracts their phase. Background phase from field inhomogeneity cancels, leaving a phase difference that measures velocity in each voxel.
Velocity encoding (VENC)
The scanner scales the gradients so that a chosen velocity, the VENC, produces a phase of exactly ±180° (±π):
phase = π × v / VENC
Velocities above the VENC wrap around and are shown as flow in the opposite direction. This is velocity aliasing:
import math
def measured_velocity(v, venc):
phase = math.pi * v / venc
wrapped = (phase + math.pi) % (2 * math.pi) - math.pi # phase is only known within ±π
return venc * wrapped / math.pi
print(measured_velocity(80, 100)) # 80.0 (correct)
print(measured_velocity(120, 100)) # -80.0 (aliased)
A VENC that is too low causes aliasing; one that is far too high makes slow flow noisy because the phase changes are tiny. A common approach is to set the VENC a little above the highest velocity expected, and to repeat with a higher VENC if aliasing appears.
| Application | Typical VENC range | Note |
|---|---|---|
| Aortic and pulmonary flow | ≈ 150–200 cm/s | Higher in valve stenosis |
| Venous flow | ≈ 30–50 cm/s | |
| CSF flow at the cerebral aqueduct | ≈ 5–20 cm/s | Cardiac-gated |
These ranges are typical starting points; local protocols and the expected pathology decide the value.
Clinical uses
- Cardiac: stroke volume, regurgitant fraction, shunt ratio (Qp:Qs) and peak velocity across stenotic valves.
- Vascular: phase-contrast angiography without contrast, and flow in dissections and collaterals.
- CSF: flow through the aqueduct and at the foramen magnum.
- 4D flow: time-resolved 3D phase contrast with velocity encoded in all three directions, used to visualise complex flow patterns.
Practical tips
- Place the imaging plane perpendicular to the vessel for through-plane measurements.
- Use ECG gating for pulsatile flow, and enough cardiac phases to capture the peak.
- Check for aliasing on the phase images before measuring.
- Background phase offsets from eddy currents can bias flow volumes; many scanners offer correction.
References
- Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. RadioGraphics 2002;22(3):651–671.
- Markl M, Frydrychowicz A, Kozerke S, Hope M, Wieben O. 4D flow MRI. J Magn Reson Imaging 2012;36(5):1015–1036.
Related articles
- Circle of Willis on MRI and time-of-flight MRA
- MRI gradients and spatial encoding
- MRI artefacts: a practical guide
Last reviewed: 7 November 2026. Educational content for researchers, clinicians and students. Values are typical examples; local protocols, scanner software and manufacturer guidance take priority.