MRI Safety

MRI implant and device screening: reading MR Conditional labels step by step

Learning objectives

  • Explain the three ASTM F2503 terms and what each one allows
  • Follow a structured screening pathway from referral to scan room entry
  • Read an MR Conditional label and check each condition against your scanner
  • Understand why projectile risk is highest at the bore entrance
ASTM F2503 MR Safe, MR Conditional and MR Unsafe labels
The three ASTM F2503 marking labels: MR Safe, MR Conditional and MR Unsafe. Label graphics: Nelissenjules, CC BY-SA 4.0, via Wikimedia Commons; arranged by MRISpin.

Why screening is the most important MRI safety step

A superconducting MRI magnet is always on. There is no switch that makes the scanner room safe for a ferromagnetic object, a device that has not been checked, or an implant used outside its labelled conditions. Almost every serious MRI incident starts with something or someone entering Zone IV without being screened properly. Screening is therefore not a form to sign but a process with several independent checks.

MRI accident: floor polishing machine pulled into a 1.5 T scanner bore
A floor-polishing machine pulled into the bore of a 1.5 T scanner. It could only be removed by ramping down the magnetic field. Image: Xksev, CC BY-SA 4.0, via Wikimedia Commons.

The photograph above shows what a single unscreened object can do. A floor-polishing machine was taken into the scanner room and pulled into the bore. It could only be removed by ramping down the magnetic field, which takes the scanner out of service. With a patient on the table, the same event could have caused fatal injury.

The three ASTM F2503 terms

TermWhat it meansIn practice
MR SafeThe item poses no known hazards in any MR environment. It is made of non-conducting, non-metallic and non-magnetic materials.Plastic items such as some wedges, pads and phantoms. The label must be on the item; a plastic look is not enough.
MR ConditionalThe item has been shown to pose no known hazards in a specified MR environment under specified conditions of use.Most modern implants and much of the equipment used in the scanner room. Every condition on the label must be met.
MR UnsafeThe item poses unacceptable risks in all MR environments.Ferromagnetic tools, oxygen cylinders, many older devices. It must never enter Zone IV.

Older labels such as “MR compatible” or “non-magnetic” are not part of the current terminology. Treat items carrying only those words as unlabelled and check them with your MR safety expert.

Screening pathway from referral to scan room

  1. Referral. The request should state implants, devices, previous surgery and relevant history so that problems are found before the appointment day.
  2. Written questionnaire. The patient (or a responsible person) completes the department’s screening form: implants, pacemakers or defibrillators, neurostimulators, cochlear implants, shunts, metal fragments, eye injuries with metal, tattoos, drug patches, pregnancy, previous MRI.
  3. Identify each implant exactly. Obtain the manufacturer, model and, where needed, the lead or component combination from the implant card, operation notes or the implanting team. “A hip replacement” is not enough for an MR Conditional decision.
  4. Find the current labelling. Use the manufacturer’s current instructions for use or MRI information for that exact model. Conditions can change between versions.
  5. Compare the label with your scanner. Field strength, spatial field gradient at the positions the patient will pass, RF limits, gradient limits and any required scan mode (see the next section).
  6. Decide and document. Unclear cases go to the MR safety expert and the responsible doctor under local policy. Record the device, the conditions used and who approved the scan.
  7. Verbal re-check at Zone III. Go through the form again with the patient immediately before entry. Patients often remember things only when asked a second time.
  8. Change and metal check. Patients change into a gown and remove all loose metal, cards, hearing aids and phones. A ferromagnetic detection system, where available, is an extra check and does not replace screening.
  9. Set up the scan to the conditions. Enter the required RF and gradient limits on the console before the first sequence, and keep them for the whole examination.

Reading an MR Conditional label

An MR Conditional label lists the conditions under which the device was tested. The exact wording differs between manufacturers, but most labels contain the items below. Every one of them has to be satisfied, not only the field strength.

ConditionWhat to check
Static field strengthOften 1.5 T only, or 1.5 T and 3 T. A device conditional at 1.5 T is not automatically conditional at 3 T.
Maximum spatial field gradientGiven in T/m or gauss/cm. Compare it with the field-gradient map from your scanner manufacturer for the positions the patient will pass through.
RF exposureA whole-body or head SAR limit, a B1+rms limit in µT, and sometimes a required operating mode such as Normal Operating Mode.
Transmit coilSome devices allow only a transmit/receive head or extremity coil, or exclude certain body regions from the transmit field.
Gradient limitsSome neurostimulators and cardiac devices limit the gradient slew rate.
Scan timeMaximum active scan time per session, sometimes with cooling periods.
Device settingsActive implants may need to be switched to an MRI mode and checked before and after the scan by a trained person.
Patient conditionsFor example, no abandoned or fractured leads, or a minimum time after implantation.

Labels usually also describe the expected temperature rise and the size of the image artefact. These help with planning but are not permissions.

Why the bore entrance is the most dangerous place

Two different forces act on ferromagnetic objects. Torque tries to align an elongated object with the field and is strongest where the field itself is strongest, at the centre of the magnet. Translational (attractive) force pulls the object towards the magnet and depends on the spatial gradient of the field. The spatial gradient is highest around the bore entrance and the front of the magnet, not at the centre. This is why the spatial field gradient on a device label must be compared with the scanner’s field map rather than with the field strength alone, and why objects can accelerate quickly over a short distance near the scanner.

Special situations

  • Possible metal in the orbits. Patients with a history of grinding, welding or eye injury involving metal need further assessment under local policy before entry, for example orbital radiographs if the history cannot exclude a fragment.
  • Unconscious or unscreenable patients. Use records, relatives, the referring team and imaging such as previous radiographs. A risk–benefit decision is made by the responsible doctor with the MR safety team.
  • Equipment brought by other teams. Monitors, pumps, ventilators, trolleys and wheelchairs need an MR Safe or MR Conditional label for your scanner; otherwise the patient is transferred to MR-labelled equipment in Zone III.
  • Emergency in the scanner room. The patient is moved out of Zone IV for resuscitation. Emergency teams do not rush into the room with their own equipment. If a person is pinned by an object and life is at risk, the magnet may need to be quenched according to local procedure.

Frequently asked questions

If a patient had an MRI before with the same implant, can they be scanned again?

A previous scan is useful information but not proof. The previous scanner, field strength, coil or settings may have been different, and labelling can change. Each examination still needs the device checked against the current conditions.

Is a hand-held magnet or detector enough to clear an object?

No. A weak attraction to a hand magnet does not predict the force near a clinical scanner, and a non-ferromagnetic item can still heat or interfere with the image. Decisions are based on labelling and expert advice.

References and further reading

  • Kanal E, Barkovich AJ, Bell C, et al. ACR guidance document on MR safe practices: 2013. J Magn Reson Imaging 2013;37(3):501–530. doi:10.1002/jmri.24011
  • Medicines and Healthcare products Regulatory Agency (MHRA). Magnetic resonance imaging equipment in clinical use: safety guidelines. GOV.UK (current version).
  • ASTM F2503. Standard practice for marking medical devices and other items for safety in the magnetic resonance environment. ASTM International (current edition).
  • Panych LP, Madore B. The physics of MRI safety. J Magn Reson Imaging 2018;47(1):28–43. doi:10.1002/jmri.25761

Related articles

Last reviewed: 6 November 2026. Educational content only. Example parameters are illustrative and vary by scanner, vendor and local protocol. It does not replace your institution’s policies, the manufacturer’s instructions, or the judgement of your MR safety expert and reporting radiologist.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles

Back to top button