MRI Safety

Pacemakers, ICDs and MRI: MR Conditional systems and the safe workflow

Learning objectives

  • Explain the risks of MRI for patients with pacemakers and implantable cardioverter-defibrillators (ICDs)
  • Check whether a device is an MR Conditional system
  • Follow the steps before, during and after the scan
  • Know the options for non-conditional (legacy) devices
Chest X-ray of a dual-chamber pacemaker with right atrial and right ventricular leads
Dual-chamber pacemaker in an 89-year-old man, placed through the left subclavian vein, with leads in the right atrium and right ventricle. Image: Stephanie C Torres et al., CC BY 4.0, via Wikimedia Commons.

Why cardiac devices need special care

Cardiac implantable electronic devices (CIEDs) include pacemakers, ICDs and cardiac resynchronisation therapy (CRT) devices. The MRI environment can affect them in several ways:

  • Heating at the lead tips from the RF field, which could damage the heart muscle where the lead touches it.
  • Unwanted pacing or inhibition of pacing caused by the switching gradients and RF pulses.
  • Inappropriate shocks from an ICD that mistakes interference for a dangerous rhythm.
  • Device reset or battery drain, and force or torque on the generator in the static field.

MR Conditional means the whole system

A device is MR Conditional only when the generator and every lead are an approved combination from the manufacturer. A conditional generator with an older lead, an abandoned lead, a lead adaptor or an epicardial lead may not meet the conditions. The labelling usually specifies:

  • Allowed field strength (often 1.5 T, sometimes 3 T)
  • SAR and gradient slew rate limits
  • Any scan regions that are excluded
  • A minimum time since implantation (often about 6 weeks)
  • That the device must be in its MRI mode during the scan
Chest X-ray of a single-chamber (VVI) pacemaker
Single-chamber (VVI) pacemaker on a chest X-ray. Image: Th. Zimmermann, CC BY-SA 3.0 DE, via Wikimedia Commons.

The workflow

  1. Identify the system: device card, implant records or chest X-ray, checked against the manufacturer’s list of conditional combinations.
  2. Cardiology or physiology review: check the device and leads, battery and whether the patient depends on pacing.
  3. Program MRI mode: asynchronous pacing for pacing-dependent patients, or pacing off for those who are not; tachycardia therapies off on ICDs.
  4. Monitor during the scan: ECG and pulse oximetry, with staff trained in advanced life support and a defibrillator available outside the magnet room.
  5. Scan within the conditions: field strength, SAR mode and any other limits on the label.
  6. Restore settings after the scan: reprogram the device and check it before the patient leaves.

If the patient has a cardiac arrest, they are moved out of the magnet room to Zone III for resuscitation; defibrillators and most resuscitation equipment are not safe in Zone IV. See MRI safety zones explained.

X-ray of a cardiac resynchronisation therapy device with three leads
Cardiac resynchronisation therapy device with a right atrial lead, a right ventricular lead and a coronary sinus lead. Image: Gregory Marcus, MD, CC BY 3.0, via Wikimedia Commons.

Legacy (non-conditional) devices

Large studies, including the MagnaSafe registry, have shown that patients with non-conditional pacemakers and ICDs can be scanned at 1.5 T with a low rate of problems when a strict protocol is followed. The 2017 Heart Rhythm Society expert consensus statement supports scanning such patients when the benefit is clear and the facility has a defined process, with the same programming and monitoring steps as above. Local policy decides whether and how a department offers this.

Key points

  • Check the whole system: generator plus every lead.
  • Program MRI mode before and restore settings after.
  • Monitor throughout and resuscitate outside Zone IV.

References

  • Indik JH, Gimbel JR, Abe H, et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm 2017;14(7):e97–e153.
  • Russo RJ, Costa HS, Silva PD, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med 2017;376(8):755–764.

Related articles

Last reviewed: 7 November 2026. Educational content only. Every site must follow its own MR safety policy, the device manufacturer’s MR Conditional instructions and national guidance, under the direction of its MR safety expert and responsible staff.

MRISpin Team

MRISpin team combines decades of experience in MRI research, clinical practice, and science communication. Meet the people driving MRISpin’s mission.

Related Articles

Back to top button